Sample sets
What belongs in the sample set
WHO asks for a defined sample set: checked vial order, a blank where carry-over matters, SST inside the set, and bracketing standards at set intervals.
WHO’s good chromatography practices guideline (2020) has a short section on the sample set. It reads like a checklist.
The WHO checklist
- Define the sample set, and check the vials are in the right order before you start.
- Put a blank in where carry-over or interference is a risk. Blank peaks have to stay within set limits.
- The SST is part of the sample set. It has to pass before samples, and stay passing.
- Bracketing standards go in at set intervals, and the procedure says how many.
- Acceptance criteria exist for SST, bracketing standards, relative retention and peak shape.
A typical QC order
This is a common pattern, not a rule; your method decides:
| Row | What | Why |
|---|---|---|
| 1 | Blank | Shows the system and diluent are clean |
| 2 | SST standard, 5 or 6 injections | Repeatability, plates, tailing, resolution |
| 3 | Check standard | Confirms the standard preparation |
| 4-9 | Samples | |
| 10 | Bracketing standard | Shows the system held steady |
| … | Samples, brackets | Repeat at the set interval |
| End | Bracketing standard | Closes the last bracket |
In Empower
The Describe Standard Bracket page of the wizard sets the vials per group, injections per vial, volume, run time and method set. Each row has a function such as Inject Standards or Inject Samples, which tells Empower which injections build the calibration.
The SST calculations themselves are explained on the system suitability page in Inside the HPLC.
The numbers
- Before the run
- Sample set defined; vials checked for the right injection order 1
- Blank
- Included where carry-over or interference matters; blank peaks within set limits 1
- SST
- Part of the sample set; passes before samples and throughout 1
- Bracketing
- Standards at defined intervals; their number defined; criteria checked 1
- Criteria include
- SST, bracketing standards, relative retention, peak shape 1
- Standard bracket in the wizard
- Vials per group, injections per vial, volume, run time, method set 2
- SST failure option
- Interactive system suitability can be set to Continue on Fault 2
What goes wrong
- SST passed at the start but nothing checks the end of the runCheck: WHO: SST has to meet its criteria before samples and throughout. Bracketing standards with criteria are how you show 'throughout'. 1
- The sample set ran on after the SST failedCheck: Check the sample set method's suitability setting. The Getting Started example uses Continue on Fault, which is fine for training but not what most QC procedures want. 2
Check yourself
Tap an answer to see if you are right and why.
Sources
- World Health Organization, 2020, WHO Good Chromatography Practices, TRS 1025, Annex 4. Sections 11.7-11.13, pp. 123-124.
- Waters, 2017, Empower 3 Getting Started Guide (715005480). Describe Standard Bracket page, p. 65; runtime options, pp. 67-68.