Results and calibration

Calibration curve fits and weighting

Empower offers single-level, multilevel and forced-through-zero fits. Weighting works on some fit types only. Average By decides whether replicate standards are averaged first.

The fit type lives in the processing method, and it should match what the method was validated with.

The choices

Group Fits
Single level (one standard) Linear through zero, response factor
Multilevel Linear, inverse linear, log-log linear, quadratic, cubic, 4th order, 5th order, power
Multilevel through zero Linear, quadratic, cubic, 4th and 5th order, response factor
Other Point-to-point, cubic spline

The default is Linear.

Average By

This setting decides what to do with replicate injections of the same standard. None (the default) fits every injection as its own point. Amount averages the responses at each level first, then fits. Waters’ example: the caffeine result barely moved, but r and r² changed slightly.

Forcing through zero

Waters’ example on the same data: Linear gave 0.039920 mg/mL; Linear through zero gave 0.040102 mg/mL. Small, but it is a different number, so it has to be the validated choice.

Weighting

Weighting gives more say to the levels with less scatter. Empower allows it only on linear, quadratic, cubic, 4th and 5th order fits. Waters’ advice is practical: plot the standard deviation at each level against the amount, and choose the weighting that matches how the scatter grows.

Point-to-point, a warning

It joins the points with straight lines, so r is always 1 and the standard error 0. A perfect r here proves nothing.

⚠ The Theory Guide’s table writes the forced-through-zero quadratic as y = B + Cx², which looks like a typo for y = Bx + Cx².

The numbers

Single level
Linear through zero, or response factor 1
Multilevel
Linear, inverse linear, log-log, quadratic, cubic, 4th and 5th order, power 1
Multilevel through zero
Linear, quadratic, cubic, 4th and 5th order, response factor 1
Also
Point-to-point, cubic spline 1
Default fit
Linear 2
Average By
Default None; Amount averages the replicate standards at each level 2
Weighting allowed on
Linear, quadratic, cubic, 4th and 5th order only 1
Choosing a weighting
Plot the standard deviation of each level against amount 1
Point-to-point
Passes through every point: r = 1, standard error 0, no coefficients 1

What goes wrong

  • A point-to-point curve gives r = 1.000Check: Waters: it always does, because the curve goes through every point. It says nothing about linearity. 1
  • Peak code Q10 on a sampleCheck: Waters: the sample's response is below the lowest standard. The result is outside the calibrated range. 2
  • Results moved after the fit was changed to Linear through zeroCheck: Waters' example: caffeine went from 0.039920 to 0.040102 mg/mL. The fit type is part of the validated method; check the method version. 2

Check yourself

Tap an answer to see if you are right and why.

On which fit types can Empower apply weighting?

Sources

  1. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Section 4.4, pp. 82-83; 4.4.3.1 point-to-point, p. 88; 4.4.5, pp. 96-97.
  2. Waters Help Center, Empower Tip #269: Calibration curve parameters in the processing method that affect amount/concentration. Average By and Fit.