Results and calibration
Calibration curve fits and weighting
Empower offers single-level, multilevel and forced-through-zero fits. Weighting works on some fit types only. Average By decides whether replicate standards are averaged first.
The fit type lives in the processing method, and it should match what the method was validated with.
The choices
| Group | Fits |
|---|---|
| Single level (one standard) | Linear through zero, response factor |
| Multilevel | Linear, inverse linear, log-log linear, quadratic, cubic, 4th order, 5th order, power |
| Multilevel through zero | Linear, quadratic, cubic, 4th and 5th order, response factor |
| Other | Point-to-point, cubic spline |
The default is Linear.
Average By
This setting decides what to do with replicate injections of the same standard. None (the default) fits every injection as its own point. Amount averages the responses at each level first, then fits. Waters’ example: the caffeine result barely moved, but r and r² changed slightly.
Forcing through zero
Waters’ example on the same data: Linear gave 0.039920 mg/mL; Linear through zero gave 0.040102 mg/mL. Small, but it is a different number, so it has to be the validated choice.
Weighting
Weighting gives more say to the levels with less scatter. Empower allows it only on linear, quadratic, cubic, 4th and 5th order fits. Waters’ advice is practical: plot the standard deviation at each level against the amount, and choose the weighting that matches how the scatter grows.
Point-to-point, a warning
It joins the points with straight lines, so r is always 1 and the standard error 0. A perfect r here proves nothing.
⚠ The Theory Guide’s table writes the forced-through-zero quadratic as y = B + Cx², which looks like a typo for y = Bx + Cx².
The numbers
- Single level
- Linear through zero, or response factor 1
- Multilevel
- Linear, inverse linear, log-log, quadratic, cubic, 4th and 5th order, power 1
- Multilevel through zero
- Linear, quadratic, cubic, 4th and 5th order, response factor 1
- Also
- Point-to-point, cubic spline 1
- Default fit
- Linear 2
- Average By
- Default None; Amount averages the replicate standards at each level 2
- Weighting allowed on
- Linear, quadratic, cubic, 4th and 5th order only 1
- Choosing a weighting
- Plot the standard deviation of each level against amount 1
- Point-to-point
- Passes through every point: r = 1, standard error 0, no coefficients 1
What goes wrong
- A point-to-point curve gives r = 1.000Check: Waters: it always does, because the curve goes through every point. It says nothing about linearity. 1
- Peak code Q10 on a sampleCheck: Waters: the sample's response is below the lowest standard. The result is outside the calibrated range. 2
- Results moved after the fit was changed to Linear through zeroCheck: Waters' example: caffeine went from 0.039920 to 0.040102 mg/mL. The fit type is part of the validated method; check the method version. 2
Check yourself
Tap an answer to see if you are right and why.
Sources
- Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Section 4.4, pp. 82-83; 4.4.3.1 point-to-point, p. 88; 4.4.5, pp. 96-97.
- Waters Help Center, Empower Tip #269: Calibration curve parameters in the processing method that affect amount/concentration. Average By and Fit.
New modules by email
I am adding the Empower pages area by area, together with the live group Empower 3 for Beginners. Leave your email and I will send you each one when it is ready.