nalam.ca Learn
Sign in Join a group

Click a box to open the area, or click any item inside it to go straight to that page. The bar at the top opens Project and access.

Review and audit trail

The printout is not the raw data

A printed or PDF chromatogram is a static copy. Review and sign on the electronic record, where the baselines, versions and audit trail are.

Many labs still sign a printed report. That is fine as a summary. It is not the record.

What the regulators say

MHRA: when data are captured electronically, the paper copy is not raw data, and keeping printouts in place of the electronic record isn’t appropriate. FDA: a chromatogram is a dynamic record, because a moved baseline changes the peaks. Waters says the same about Empower reports: a printout can’t show whether the baselines were right.

What the printout hides

Waters’ own example: a summary report showed a sample passing. In Empower, the channel had three results, each with the baseline moved a little more, and only the passing third result was on the report. Nothing on paper showed it.

Signing in Empower instead

Electronic signatures are part of the Empower base licence. To use them properly, Waters lists:

  • Enforce Unique User Account Names, so a signature is one real person.
  • Report methods created with Allow this method to be used for sign off.
  • Separate Sign Off 1 (analyst) and Sign Off 2 (approver) privileges, or the policy that needs two different users.
  • Optional policies: clear the password after signing, lock channels after Sign Off 2 so no more results can be made, and save the signed report as a PDF.

What a reviewer does

Open the result in Review. Zoom into the baselines of the reported peaks, read the Int Type letters, check the processing method version, look at the channel’s other results, then sign. The printout is what comes out after.

The numbers

Raw data (MHRA)
Where data are captured electronically, paper copies are not raw data 1
Keeping printouts instead (MHRA)
Not appropriate if the system can store the dynamic data 1
Dynamic record (FDA)
A chromatographic record is dynamic: changing the baseline changes the peaks 2
Printed report (Waters)
Static; the review of baselines is lost 3
E-signatures (Waters)
Part of the base licence; need report methods allowed for sign off 4
After sign off
Policies can lock channels after Sign Off 2 and save the report as a PDF 5

What goes wrong

  • The summary report passes, the data don'tCheck: Waters' own example: three results on one channel, only the passing third one reported. Review in Empower with View As > Results, not on the printout. 4
  • Analysts can approve their own resultsCheck: Waters: give analysts Sign Off 1 only and approvers Sign Off 2 only, or set the policy requiring two different users. 4

Check yourself

Tap an answer to see if you are right and why.

Data are captured electronically in Empower. Is the signed printout the raw data?

Sources

  1. UK MHRA, March 2018, MHRA 'GXP' Data Integrity Guidance and Definitions, Revision 1. Sections 6.2, 6.7.
  2. US FDA, December 2018, Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry. Definitions, p. 5.
  3. Waters, 2017, Empower Software Audit Trails and Logs: a guide to the different locations of audit trails in Empower (white paper). p. 7.
  4. Waters Help Center, Empower Tip #171: How to use electronic signatures to approve results. Electronic signatures; review thoroughly before you sign.
  5. Waters Help Center, Empower Tip #385: Result sign off system policies (part 1). Result sign off policies.

New modules by email

I am adding the Empower pages area by area, together with the live group Empower 3 for Beginners. Leave your email and I will send you each one when it is ready.

My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy