Module
Processing and integration
Processing finds the peaks, draws the baselines and measures them. In Empower that is ApexTrack or Traditional integration, plus events. Most review questions start here.
Processing is where the same raw data can give different numbers. So it is where analysts spend time and where reviewers look first.
Empower has two integration algorithms. Traditional finds a peak where the signal lifts off the baseline. ApexTrack finds a peak at its apex, using the curvature (the second derivative), so a sloping baseline doesn’t stop it finding the peak. In ApexTrack, finding the peak and drawing the baseline are two separate steps with separate settings.
The pages below go through the settings in the order they act: peak width and threshold (is it a peak?), liftoff and touchdown (where is the baseline?), the integration type codes (what did the software do?), drop line or skim, and manual integration.
The general version, for any data system, is on the integration page in Inside the HPLC.
The numbers
Module check
5 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.
Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.
Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →
Go deeper
- Peak Width and Detection Threshold In ApexTrack these two settings decide which peaks exist. Leave them blank and Empower picks values for every chromatogram, and it can pick badly.
- Liftoff % and Touchdown % In ApexTrack these set where each baseline starts and ends. Higher values start and end the peak further up its sides. Touchdown is the one that matters for tailing peaks.
- Int Type: the two letters on every peak Each peak gets two letters that describe its start and end. Capitals mean the processing method did it. Lowercase means someone moved it by hand.
- Drop line or skim For a small impurity on the tail of a big peak, a drop line or a skim can change the impurity result by more than 100%. The lab needs a written rule for which one.
- Manual integration Moving a baseline by hand is allowed in exceptional cases, by trained users, with a reason. Processing again and again until a result passes is what inspectors look for.
Sources
- Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Section 2.1.2, p. 12; section 2.5, p. 41.