Test injections

Waters' Getting Started guide makes a single injection called Test Injection. WHO says injections outside the sample set are not recommended. Know both.

In the Getting Started guide, Waters teaches a single injection and names it Test Injection. It is a training exercise in a training project, and it makes sense there.

Why it is a problem in GMP

WHO’s guideline says trial or system check injections that are not part of the sample set are not recommended. If a lab really needs them, a written procedure describes when, normally with standard solution only (biologics may justify more), and the electronic records are kept with the sample set’s results.

The reason is simple. A sample injected “just to check” and then not reported looks exactly like testing until it passes.

What to do instead

  • Watch the baseline with Equilibrate System/Monitor Baseline. It runs the instrument method without an injection.
  • Put the system check where it belongs: the SST injections at the start of the sample set.
  • If a check standard is allowed before the set, follow the procedure, keep the data, and report it with the set.

What a reviewer checks

In the project, list the injections by date and time. Look for injections on the same system before the sample set started, especially ones with a sample name, or in a different project.

The interview question

“Have you ever made a test injection?” A good answer explains both sides: Waters uses one for training, and WHO’s guideline says why a GMP lab shouldn’t.

The numbers

Waters training example
A single injection named Test Injection, before building a sample set 1
WHO position
Trial or system check injections outside the sample set are not recommended 2
If a lab does them (WHO)
An authorized procedure; normally standard solutions only, unless justified (e.g. biologics) 2
Records (WHO)
Saved and kept with the results of the sample set 2
Equilibration
Equilibrate System/Monitor Baseline uses an instrument method without injecting 1

What goes wrong

  • Several single injections of the sample before the official sample setCheck: WHO: the pattern inspectors look for. Injections outside the sample set should not happen; if a procedure allows them they are standards, and their records stay with the set. 2
  • The system needs to be checked before the runCheck: Use Equilibrate System/Monitor Baseline to watch the baseline without injecting, and put the SST in the sample set itself. 1

Check yourself

Tap an answer to see if you are right and why.

What does WHO say about trial injections outside the sample set?

Sources

  1. Waters, 2017, Empower 3 Getting Started Guide (715005480). Section 4.3, Acquiring data, pp. 40-41.
  2. World Health Organization, 2020, WHO Good Chromatography Practices, TRS 1025, Annex 4. Section 11.9, p. 123.