Methods and method sets

Instrument method and processing method

The instrument method runs the hardware and can't change the data after the run. The processing method turns the signal into numbers and can be changed later, which is why reviewers watch it.

Instrument method

It holds everything the hardware does: flow and gradient, column temperature, injection settings, wavelength, and the detector’s sampling rate and time constant. Once the run is done, these are baked into the raw data. A wrong wavelength can’t be fixed by reprocessing.

Processing method

It works on the stored data, so it can be changed later. That is why it gets so much attention in an audit. It holds:

Tab or area What it decides
Integration Algorithm (Traditional or ApexTrack), peak width, threshold, liftoff and touchdown, timed events
Components Names, retention times and windows, which peaks get an amount
Calibration Fit type, weighting, Average By
Default amounts Standard amounts and purity, if not entered in the sample set
Suitability, impurities, noise Optional calculations

ApexTrack has to be switched on by system policy and in the project before you can choose it. The Processing and integration area explains the two algorithms.

Making a processing method

Waters’ tip for ApexTrack: open a chromatogram in Review with the Processing Method layout, create a new processing method, pick ApexTrack, set the start and end time, and click Integrate. Empower shows the peak width and threshold it chose. You can leave them automatic, or right-click and Copy to Processing Method to fix them. For a validated method, fixed values are easier to defend.

Which version made this result?

Every save makes a new version, and each result stores the ID of the version that made it. Waters’ tip: right-click a result, View As Processing Methods, then use Method Properties and Differences to see what changed between versions.

The numbers

Instrument method controls
Pumps, solvent managers, detectors, autosamplers, sample managers, GCs 1
Detector settings in it
Sampling rate and time constant 3
Processing method defines
Integration, calibration, component identification, allowed ranges, noise and drift 1
Integration algorithm
Traditional (finds peak start) or ApexTrack (finds peak apex) 1
ApexTrack availability
Enabled by system policy and project properties, then chosen per processing method 2
Version used
Empower records which processing method version made each result 4

What goes wrong

  • Two results for the same injection disagreeCheck: Waters: each result stores the processing method ID and version. Right-click the result, View As Processing Methods, then compare the versions with Differences. 4
  • Wrong wavelength or data rate found after the runCheck: These are in the instrument method and are part of the raw data. Reprocessing can't fix them; the run has to be repeated under your procedure. 3

Check yourself

Tap an answer to see if you are right and why.

Which method holds the integration and calibration settings?

Sources

  1. Waters, 2017, Empower 3 Getting Started Guide (715005480). Table 1-1, Types of methods, p. 8.
  2. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Section 2.1.2.1, p. 13.
  3. Waters Help Center, Empower Tip #207: Setting the sampling rate and time constant. Sampling rate and time constant in the instrument method.
  4. Waters Help Center, Empower Tip #328: Working with multiple versions of a processing method. Processing method versions and results.