Processing and integration

Manual integration

Moving a baseline by hand is allowed in exceptional cases, by trained users, with a reason. Processing again and again until a result passes is what inspectors look for.

Sometimes the method’s settings can’t integrate one peak properly, and an analyst moves the baseline by hand. Regulators don’t forbid it. They put fences around it.

The rules

WHO: setting the baseline by hand only in exceptional cases, only by trained, experienced users who have that privilege, following an authorized procedure, with the authorization and the reason recorded.

MHRA: if a result was processed again and again with changed settings, that has to be visible.

A real pattern, from Waters’ own tip

Waters describes a case a reviewer could find in Empower:

  1. The summary report showed a sample passing.
  2. In the project, the channel had three results, and only the third was reported.
  3. Result 1: the API was below the label claim limit. Result 2: just below. Result 3: just above.
  4. Each time the baseline had been moved by hand. The first baseline, from the approved method, was fine. No reason was recorded.

Waters calls it most likely deliberate manipulation, and it was invisible on paper.

How to find it

  • Peaks tab: filter for lowercase Int Type letters.
  • Channels tab: right-click the injection and choose View As > Results. More than one result means it was processed more than once.
  • Result Audit Viewer: Result Differences shows what changed between a result and the ones it superseded.
  • Note Waters’ caution: processing the whole sample set again makes a new result set, not a “superseded” result. So check the channel, not just the flag.

The manual integration page in Inside the HPLC covers the difference between manual integration and changing the method.

The numbers

When (WHO)
Only in exceptional cases 1
Who (WHO)
Trained, experienced users with the privilege 1
Records (WHO)
Authorized procedure; authorization and justification recorded 1
Reprocessing (MHRA)
Repeated processing with changed parameters must be visible 2
In Empower
Lowercase Int Type letters 3
Superseded results
Changing a result in a result set and saving makes a superseded result 5

What goes wrong

  • Three results exist for one injection; only the last one is reportedCheck: Waters' own example: the first two failed, the third passed after baseline changes with no reason recorded. Open the channel's results with View As > Results and compare them. 4
  • Reprocessing the whole sample set left no superseded resultCheck: Waters: processing a sample set again makes a second result set, not a superseded result. Check the channel for all its results, not just the superseded flag. 5

Check yourself

Tap an answer to see if you are right and why.

What does MHRA say about processing a result again and again with changed settings?

Sources

  1. World Health Organization, 2020, WHO Good Chromatography Practices, TRS 1025, Annex 4. Sections 13.5-13.6, p. 125.
  2. UK MHRA, March 2018, MHRA 'GXP' Data Integrity Guidance and Definitions, Revision 1. Section 6.9, data processing.
  3. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Section 2.1.5, p. 15.
  4. Waters Help Center, Empower Tip #171: How to use electronic signatures to approve results. Review thoroughly before you sign.
  5. Waters Help Center, Empower Tip #81: Using the Result Differences table in the Result Audit Viewer. Result Differences; final notes.