Sample sets
Changing sample details after the run
Alter Sample fixes names, weights, dilutions and standard amounts after acquisition. It can't change what physically happened, and every change leaves a flag and a history.
Typing mistakes happen: a wrong weight, a wrong dilution, a standard amount from last month’s lot. Alter Sample is where they get fixed after the run.
What you can and can’t change
Waters draws a clear line. You can change what you entered: sample name, weight, dilution, component amounts and purity. You can’t change what the instrument did: vial position, injection volume.
How to do it
- In the project, open the Sample Sets tab, right-click the sample set and choose Alter Sample.
- Change the field. For standards, click Amount to open the Component Editor and enter each level.
- Save. With full audit trail on, Empower asks for a reason.
- Reprocess. The old results stay; the new processing makes new results.
The trail it leaves
- An Altered check mark on the sample row.
- View Sample History (right-click the row) shows each field’s old and new value. Waters’ example: Purity changed from empty to 99.9%.
- View Sample Set Changes shows changes to the set as a whole.
- The project audit trail records the change.
What a reviewer checks
Filter for altered samples first. For each one: what changed, who changed it, when (before or after the first result?), the reason, and the supporting record (balance printout, standard certificate). A weight changed after an OOS result, with no balance record, is the case inspectors write up.
The numbers
- Where
- Browse Project > Sample Sets tab > right-click > Alter Sample 1
- Can change
- Sample name, sample weight, dilution, component amounts 1
- Can't change
- What was used to collect the data, e.g. vial number and injection volume 1
- Altered field
- A check mark shows a sample identifier was changed 2
- See the change
- Right-click > View Sample History: old and new values 2
- Whole set
- Sample Set Information > View Sample Set Changes 3
- After a change
- Altered samples are permanently flagged; old results stay, reprocessing makes new ones 5
What goes wrong
- A sample weight was typed wrong and the result is out of specificationCheck: Alter Sample can correct it. The sample is flagged and both values are kept, so the change needs a reason, the balance printout, and a reviewer. 5
- A standard's purity was changed after the runCheck: Waters: purity multiplies the standard amount used for the curve, so it moves every result. View Sample History shows when and who. 4
Check yourself
Tap an answer to see if you are right and why.
Sources
- Waters, 2017, Empower 3 Getting Started Guide (715005480). Section 7.1, Altering samples, pp. 94-100.
- Waters Help Center, Empower Tip #371: Understanding sample audit trails. Sample audit trails.
- Waters Help Center, Empower Tip #372: Understanding sample set audit trails. Sample set audit trails.
- Waters Help Center, Empower Tip #268: Sample set method parameters that affect amount/concentration. Purity.
- Waters, 2017, Empower Software Audit Trails and Logs: a guide to the different locations of audit trails in Empower (white paper). Altered samples, p. 6.
New modules by email
I am adding the Empower pages area by area, together with the live group Empower 3 for Beginners. Leave your email and I will send you each one when it is ready.
My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy