nalam.ca Learn
Sign in Join a group

Click a box to open the area, or click any item inside it to go straight to that page. The bar at the top opens Project and access.

Sample sets

Changing sample details after the run

Alter Sample fixes names, weights, dilutions and standard amounts after acquisition. It can't change what physically happened, and every change leaves a flag and a history.

Typing mistakes happen: a wrong weight, a wrong dilution, a standard amount from last month’s lot. Alter Sample is where they get fixed after the run.

What you can and can’t change

Waters draws a clear line. You can change what you entered: sample name, weight, dilution, component amounts and purity. You can’t change what the instrument did: vial position, injection volume.

How to do it

  1. In the project, open the Sample Sets tab, right-click the sample set and choose Alter Sample.
  2. Change the field. For standards, click Amount to open the Component Editor and enter each level.
  3. Save. With full audit trail on, Empower asks for a reason.
  4. Reprocess. The old results stay; the new processing makes new results.

The trail it leaves

  • An Altered check mark on the sample row.
  • View Sample History (right-click the row) shows each field’s old and new value. Waters’ example: Purity changed from empty to 99.9%.
  • View Sample Set Changes shows changes to the set as a whole.
  • The project audit trail records the change.

What a reviewer checks

Filter for altered samples first. For each one: what changed, who changed it, when (before or after the first result?), the reason, and the supporting record (balance printout, standard certificate). A weight changed after an OOS result, with no balance record, is the case inspectors write up.

The numbers

Where
Browse Project > Sample Sets tab > right-click > Alter Sample 1
Can change
Sample name, sample weight, dilution, component amounts 1
Can't change
What was used to collect the data, e.g. vial number and injection volume 1
Altered field
A check mark shows a sample identifier was changed 2
See the change
Right-click > View Sample History: old and new values 2
Whole set
Sample Set Information > View Sample Set Changes 3
After a change
Altered samples are permanently flagged; old results stay, reprocessing makes new ones 5

What goes wrong

  • A sample weight was typed wrong and the result is out of specificationCheck: Alter Sample can correct it. The sample is flagged and both values are kept, so the change needs a reason, the balance printout, and a reviewer. 5
  • A standard's purity was changed after the runCheck: Waters: purity multiplies the standard amount used for the curve, so it moves every result. View Sample History shows when and who. 4

Check yourself

Tap an answer to see if you are right and why.

Which of these can you change with Alter Sample after the run?

Sources

  1. Waters, 2017, Empower 3 Getting Started Guide (715005480). Section 7.1, Altering samples, pp. 94-100.
  2. Waters Help Center, Empower Tip #371: Understanding sample audit trails. Sample audit trails.
  3. Waters Help Center, Empower Tip #372: Understanding sample set audit trails. Sample set audit trails.
  4. Waters Help Center, Empower Tip #268: Sample set method parameters that affect amount/concentration. Purity.
  5. Waters, 2017, Empower Software Audit Trails and Logs: a guide to the different locations of audit trails in Empower (white paper). Altered samples, p. 6.

New modules by email