Module
Sample sets
The sample set method is the plan: which vials, in what order, how many injections, what volume and run time. Running it gives a sample set, the bound raw data.
Other data systems call it a sequence or a batch. Empower has two names, and the difference matters.
The sample set method is the plan you write before the run. The sample set is what you get after: the raw data, bound together, every injection and every channel.
Each row in the plan has a function, such as Inject Standards or Inject Samples. The functions tell Empower which injections are standards for calibration and which are unknowns. Rows can also equilibrate, calibrate, quantitate or summarize.
When you run it, you choose a run mode: Run only collects raw data, Run and process also processes, Run and report also prints the report.
The pages below cover templates, what WHO expects in a sample set, test injections, and what you can change after the run.
The numbers
Same part, different names
Module check
5 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.
Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.
Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →
Go deeper
- Sample set method or template A sample set method is run as it is or edited first. A template keeps the fixed rows and only asks for what changes, such as the number of samples.
- What belongs in the sample set WHO asks for a defined sample set: checked vial order, a blank where carry-over matters, SST inside the set, and bracketing standards at set intervals.
- Test injections Waters' Getting Started guide makes a single injection called Test Injection. WHO says injections outside the sample set are not recommended. Know both.
- Changing sample details after the run Alter Sample fixes names, weights, dilutions and standard amounts after acquisition. It can't change what physically happened, and every change leaves a flag and a history.
- Data rate and points across the peak Set in the instrument method, checked after the run. Waters recommends about 15 points across a peak for LC and 25 for UPLC, and Empower counts them for every peak.
Sources
- Waters, 2017, Empower 3 Getting Started Guide (715005480). Empower workflow, p. 7; runtime options, pp. 67-68.
- Thermo Fisher Scientific, 2017, Chromeleon 7.2 SR5 Quick Start Guide. Creating a sequence, p. 13.
- Shimadzu, 2017, LabSolutions Data Acquisition & Processing Theory Guide, version 5.93 (223-60090G). Batch table, p. 29.