Module
Methods and method sets
In Empower a method is a set of instructions for one job. A method set groups them so a run can acquire, process and report in one go.
Empower splits the work into methods, one per job. Waters lists them in its Getting Started guide: instrument, processing, sample set, sample set template, report, export, and the method set.
The method set is the one people mix up. It doesn’t do anything by itself; it points to the other methods. To make even one injection, you need a method set with an instrument method in it. Add a processing and report method to the same set and Empower can acquire, process and print in one run.
This page covers the instrument and processing methods, the method set, and what happens when a method changes. The sample set method has its own area.
For the same idea in other data systems, see the Inside the HPLC data system page.
The numbers
- Instrument method
- Control settings for pumps, detectors and sample managers 1
- Processing method
- Integration, calibration and component identification 1
- Report method
- The layout of a printed or electronic report 1
- Export method
- What data to export and in which format (ASCII, AIA, PDF and others) 1
- Minimum to inject
- A method set that contains an instrument method 1
Same part, different names
What goes wrong
- Someone changed the processing method and old results look different when reopenedCheck: Waters: methods are never overwritten; each save makes a new version. Check which version each result used before comparing. 2
Module check
3 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.
Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.
Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →
Go deeper
- Instrument method and processing method The instrument method runs the hardware and can't change the data after the run. The processing method turns the signal into numbers and can be changed later, which is why reviewers watch it.
- Method set A method set points to the instrument method and, if you want, the processing, report and export methods. With several channels, 3D data or derived channels, you must process with it.
- Method versions and locking Empower keeps every saved version of every method and records which version made each result. Methods can be locked, and Differences shows what changed.
Sources
- Waters, 2017, Empower 3 Getting Started Guide (715005480). Key concepts and Table 1-1, pp. 7-9.
- Waters, 2017, Empower Software Audit Trails and Logs: a guide to the different locations of audit trails in Empower (white paper). Methods, p. 5.
- Agilent Technologies, 2022, OpenLab CDS Data Analysis Reference Guide, version 2.7 (D0013815). Integration events in the processing method, p. 46; result set summary reports, p. 173.
New modules by email
I am adding the Empower pages area by area, together with the live group Empower 3 for Beginners. Leave your email and I will send you each one when it is ready.
My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy