Results and calibration

How Empower gets the amount

The curve gives an amount from the response. Purity, sample weight and dilution then adjust it. A wrong number typed in the sample set changes every result, and the chromatogram looks perfect.

Empower does the arithmetic in a fixed order. Knowing it is how you find a wrong number fast.

For the standards (building the curve)

X-value on the curve = amount entered × purity × sample weight ÷ dilution

The amount comes from the Component Editor (or the processing method’s default amounts). Purity is entered in the Component Editor. Weight and dilution come from the sample set, and default to 1.

For the samples

  1. Match each peak to a component by retention time.
  2. Read the amount off that component’s curve from the peak’s response.
  3. Divide by sample weight and multiply by dilution.

Waters’ worked example

Their tip series uses one caffeine data set and changes one thing at a time:

What changed Standard 1 X-value Sample result
Nothing (weight 1, dilution 1) 0.0400 0.100
Purity added 0.0396 0.099
Plus weight and dilution on the standards 0.0792 0.198
Plus weight and dilution on the sample 0.0792 19.824

The chromatograms are identical in every row. Only the typed numbers changed.

Without a calibration

Empower still reports percent area and percent height for each peak, against the total of all integrated peaks. That is why a missed or extra peak changes every other peak’s % area.

The numbers

Standards need
Injection volume, and component names with amounts or concentrations 1
Standard X-value
Entered amount × purity × sample weight ÷ dilution 3
Sample result
Amount from the curve ÷ sample weight × dilution 1
Defaults
Sample weight 1 and dilution 1: no effect 2
No calibration
Percent area and percent height of each peak 1
Peak matching
Sample peaks matched to components by retention time first 1

What goes wrong

  • All results in a run are off by the same factorCheck: Check sample weight and dilution in the sample set, and the standard amounts and purity in the Component Editor. Empower uses them directly. 3
  • Weight and dilution were entered for the standards by mistakeCheck: Waters: for standards, the amount is multiplied by the weight and divided by the dilution before it goes on the curve. That moves the whole curve. 3

Check yourself

Tap an answer to see if you are right and why.

A dilution of 10 was entered as 1. What happens to the sample result?

Sources

  1. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Sections 4.3-4.3.2, pp. 66-67.
  2. Waters Help Center, Empower Tip #267: How Empower calculates amount/concentration. External standard example.
  3. Waters Help Center, Empower Tip #268: Sample set method parameters that affect amount/concentration. Purity, sample weight and dilution.