Sample sets

Data rate and points across the peak

Set in the instrument method, checked after the run. Waters recommends about 15 points across a peak for LC and 25 for UPLC, and Empower counts them for every peak.

The data rate (sampling rate) is set in the detector part of the instrument method, before the run. Too low and narrow peaks are drawn from too few points, so retention time and area wobble from injection to injection. Too high and the files get big for nothing.

The rule

Waters gives it in the Theory Guide: sampling rate = N / W.

  • W is the width, in seconds, of the narrowest peak you care about.
  • N is the number of points you want across it: 15 for LC, 25 for UPLC.

Worked: a 3-second peak at 15 points needs 5 points per second. A 2-second UPLC peak at 25 points needs 12.5, so pick the next rate up.

The time constant

The time constant is a noise filter on the whole chromatogram. Waters’ tip suggests about 1 / sampling rate. Higher than that and it starts to flatten and widen the peaks.

Check it after the run

Empower counts the points for every peak: add Points Across Peak to the Peaks table in Review. Waters’ tip example: a 5.5-second peak at 10 points per second has 55 points, plenty.

For PDA data the same sampling rate also sets the number of spectra. Peak purity needs at least 12 across the peak.

The detector side of this is on the data rate page in Inside the HPLC.

The numbers

Formula
Sampling rate = N / W; W = width in seconds of the narrowest peak 1
N recommended
15 points for LC, 25 for UPLC 1
Worked example
3 s peak, 15 points: 5 points per second 1
Not available?
Pick the next higher rate 1
Where to check
Points Across Peak, on the Peaks tab in Review 1
Time constant (Waters tip)
About 1 / sampling rate; too high distorts peaks 2
PDA peak purity
At least 12 spectra from peak start to end 2

What goes wrong

  • UPLC peaks look jagged and areas repeat badlyCheck: Look at Points Across Peak in Review. Waters recommends 25 for UPLC; if it is far below, raise the sampling rate in the instrument method. 1
  • Peaks look broad and short after a detector setting changeCheck: Waters: a time constant that is too high filters the peaks too. Set it to about 1 / sampling rate. 2
  • Peak purity results are unreliableCheck: Waters: purity needs at least 12 spectra across the peak; the PDA sampling rate sets both points and spectra. 2

Check yourself

Tap an answer to see if you are right and why.

How many points across a peak does Waters recommend for UPLC?

Sources

  1. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Sections 1.2.1-1.2.3, pp. 7-8.
  2. Waters Help Center, Empower Tip #207: Setting the sampling rate and time constant. Sampling rate and time constant.

New modules by email

I am adding the Empower pages area by area, together with the live group Empower 3 for Beginners. Leave your email and I will send you each one when it is ready.

My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy