Detector

Data rate and response time

How many points per second the detector records, and how much it smooths them. Too slow and narrow peaks get flattened; too fast and the baseline gets noisy.

The detector does not draw a smooth line. It records points, and the software joins them. The data rate is how many points per second; the response time (or time constant) is how much each point is averaged with its neighbours.

A peak needs enough points to have its true shape. Waters asks for at least 15 points across the narrowest peak, and says to lower the rate if you get more than 30. Agilent says up to 20 Hz is enough for normal HPLC, and at least 40 Hz for fast columns with narrow peaks.

Smoothing is the other half. A slow time constant gives a quiet baseline, but it also rounds off narrow peaks. Agilent’s rule: set the peak width to the narrowest peak in your sample, and don’t go faster than you need, because a faster response gives more noise.

Agilent shows the effect in its handbook: the same separation gives higher efficiency as the data rate goes up. So when a method moves to a faster column, check the data rate too.

The numbers

Standard HPLC (Agilent)
Up to 20 Hz is enough 1
Fast columns (Agilent)
At least 40 Hz with a fast response time for sub-2 µm and Poroshell-type columns 1
Points across a peak (Waters)
At least 15; above 30, lower the rate 2
Agilent peak width setting
Set to the narrowest peak you expect (width at half height) 1
Waters filter at 20 Hz
Time constant 0.5 s slow, 0.2 s normal, 0.05 s fast 2
Maximum rate (Agilent 1290 Infinity II DAD)
240 Hz 1
Maximum rate (Agilent 1260 VWD)
120 Hz, single wavelength 3
Rates (Waters ACQUITY PDA)
1, 2, 5, 10, 20, 40 and 80 Hz 2

Same part, different names

Agilent
Peak width (response time), data rate 1
Waters
Sampling rate, filter time constant 2

What goes wrong

  • Peaks broader and resolution lower than expected on a fast methodCheck: Data rate too low or time constant too slow. Agilent: set the fastest value that does not cost signal-to-noise. 1
  • Baseline noisier after a method was copied to a new systemCheck: Response time set faster than needed. Agilent: a faster response time than you need gives more baseline noise. 1

Check yourself

Tap an answer to see if you are right and why.

How many data points across the narrowest peak does Waters ask for, at least?

Sources

  1. Agilent Technologies, The LC Handbook. Data rate and Max-Light box, p. 27; setting the data collection rate, p. 49; Poroshell tips, p. 89; troubleshooting, peak broadening, p. 126.
  2. Waters, ACQUITY UPLC PDA Detector Getting Started Guide (71500108703). Selecting the sampling rate, p. 1-13; digital filter settings, p. 1-16; specifications, p. A-3.
  3. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 36 (VWD G7114A), p. 52.