Detector

Refractive index detector (RI, RID)

Sees almost anything, including sugars and polymers with no UV absorbance, by measuring how the mobile phase bends light. It is very sensitive to temperature, flow pulses and mobile phase changes.

An RI detector compares how much the eluent bends light with a reference cell. Almost every compound changes the refractive index a little, so it sees things UV cannot. Agilent names carbohydrates, lipids and polymers, and it is the usual detector for GPC/SEC.

The price is that it also sees everything else. A change in temperature, a pump pulse or a change in mobile phase composition all change the refractive index. Agilent’s manual says the RID is often blamed for instability that comes from the room or the rest of the system. That is why RI methods keep one mobile phase composition (isocratic), why Agilent says not to switch the pump off between runs, why the optics are kept at a set temperature (Agilent: up to 55 °C), and why Agilent says to cover the detector and keep it out of drafts.

Give it time. KNAUER: the RI cell needs a stable temperature, and you flush it to remove air and set the zero before you start.

Two things catch people. Peaks can come out negative when the sample has a lower refractive index than the mobile phase; Dolan’s fix is to reverse the polarity. And the cell is low pressure: Agilent’s is rated 5 bar, so nothing should restrict its outlet.

The numbers

Measures
Refractive index difference between sample and reference cell (Agilent 1260 RID) 1
Sample cell (Agilent)
8 µL, 5 bar (0.5 MPa) maximum, 5 mL/min maximum flow 1
Optics temperature (Agilent)
Controlled from 5 °C above ambient to 55 °C 1
Noise (Agilent)
Below ±1.25 × 10⁻⁹ RIU; drift below 200 × 10⁻⁹ RIU/h 1
pH range (Agilent)
2.3 to 9.5 1
Before use (KNAUER)
Stable cell temperature, and flush to remove air and set the zero 4
Typical uses (Agilent)
Substances that do not absorb UV: carbohydrates, lipids, polymers; the detector of choice for GPC/SEC 2
What the signal follows (Agilent)
Temperature, pressure and solvent composition, so keep the composition stable (isocratic) 2
Between runs (Agilent)
Don't switch the pump off: recycle the solvent or lower the flow 2
Flow pulses
RI and electrochemical detectors are greatly affected by them (Agilent) 3

Same part, different names

Agilent
RID (G7162A), RID Micro (G7162B) 1

What goes wrong

  • Negative peaksCheck: The sample's refractive index is lower than the mobile phase's. Reverse the detector polarity, or use a mobile phase with a lower refractive index (Dolan). 5
  • Wavy baselineCheck: Room temperature changes. Cover the RI detector and keep it out of air currents (Agilent). 3
  • Cell damaged after a blockage or restrictor downstreamCheck: The Agilent RID cell is rated only 5 bar. Send its outlet straight to waste. 1

Check yourself

Tap an answer to see if you are right and why.

An RI detector gives negative peaks for your sample. What does Dolan suggest?

Sources

  1. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 56 (G7162A), pp. 89-90.
  2. Agilent Technologies, 2019, InfinityLab LC Series Refractive Index Detector User Manual (SD-29000641). Product description, p. 12; factors that affect refractive index, p. 17; purge and recycle valves, p. 20; detector equilibration, p. 81.
  3. Agilent Technologies, The LC Handbook. Detector types, p. 24; troubleshooting, baselines, p. 127; glossary, pulse damper, p. 160.
  4. KNAUER, 2021, Good HPLC Practice Guide (V6887). Can I start the measurement immediately after switching on the system?, p. 13.
  5. ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Negative peaks, section 3.3.