Detector

Lamp

The light source. A deuterium lamp covers the UV; a diode array detector often adds a tungsten lamp for the visible range. Lamps get weaker with hours, and noise goes up.

The lamp is a consumable. It gets weaker with every hour it is on, and a weaker lamp gives a noisier baseline.

A deuterium lamp gives continuous light in the UV, and Shimadzu puts its useful end at about 400 nm. A tungsten halogen lamp covers the visible, from about 350 nm. So a plain UV detector has one D2 lamp, and a diode array that reads up into the visible (Agilent’s DAD WR goes to 950 nm) has both.

Lamp hours are tracked. Waters warrants a lamp for 2000 hours or one year. Newer Agilent lamps carry an RFID chip with their own hours and ignitions, so the count moves with the lamp.

Turn the lamp off when the system sits idle; Waters and Shimadzu both say it saves lamp life. And let it cool for 30 minutes before you pull it out.

The student page goes into how you tell an old lamp from a dirty flow cell.

The numbers

Deuterium (D2) lamp
Continuous light in the UV, usable up to about 400 nm 1
Tungsten halogen lamp
Visible range, usable from about 350 nm 1
Agilent 1260 VWD (G7114A)
Deuterium lamp, 190 to 600 nm 2
Agilent 1260 DAD WR (G7115A)
Deuterium and tungsten lamps, 190 to 950 nm 2
Waters warranty
2000 hours or 1 year, whichever comes first (ACQUITY PDA) 4
Lamp tag (Agilent)
RFID chip on the lamp stores hours on, number of ignitions and date of last intensity test 3
When the system is idle
Turn the lamp off to save lamp life (Waters, Shimadzu) 6
Before you remove it
Let it cool for 30 minutes; never look at a lit lamp (UV) 4

Same part, different names

Agilent
Long-life deuterium lamp with RFID tag 3
Waters
Deuterium lamp (PDA) 4

What goes wrong

  • Noise or drift above what your method allows, or the lamp does not igniteCheck: Agilent's reason to change the D2 lamp. After the change, run a wavelength calibration and an intensity test. 3
  • Baseline noisy, lamp energy lowCheck: Check lamp energy and reference energy against the detector manual. Below the limit: replace the lamp (Thermo). 5

Check yourself

Tap an answer to see if you are right and why.

Which lamp gives the UV light in most HPLC UV detectors?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Shimadzu, Light Sources for Spectrophotometers (UV-Vis essential knowledge). Types of light source: halogen and deuterium lamps.
  2. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Tables 36 and 38, pp. 52 and 55.
  3. Agilent Technologies, InfinityLab LC Series Diode Array Detectors User Manual, maintenance chapter. Overview of maintenance, p. 161; replace the deuterium lamp, pp. 164-166; lamp and flow cell RFID tag, p. 178.
  4. Waters, ACQUITY UPLC PDA Detector Getting Started Guide (71500108703). Replacing the lamp, p. 5-11.
  5. Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Section 3.3.2, non-cyclic noise: detector electronics problems.
  6. Waters, ACQUITY UPLC PDA and eλ PDA Detectors: Overview and Maintenance Information (715002209). Shutting down the detector, p. 34.
  7. Shimadzu, HPLC Troubleshooting Guide. Good practice, p. 4.