Detector

Evaporative light scattering detector (ELSD)

Sprays the eluent into a gas stream, evaporates the solvent and measures the light scattered by the dry particles left. Sees compounds with no chromophore, works with gradients, but needs a fully volatile mobile phase.

The ELSD works in three steps. The eluent is sprayed into a gas stream (the nebuliser), the solvent is evaporated in a heated tube, and what is left, small dry particles of the compound, scatters light onto a photomultiplier.

Because it measures the particles and not the light the compound absorbs, it sees compounds with no chromophore. It works with gradients, unlike RI. The response changes as the solvent changes, and Agilent’s ELSD can change its gas flow during the run to make up for it.

The mobile phase has to evaporate completely. Agilent’s handbook says ELSDs need a fully volatile mobile phase, like LC-MS. Phosphate is out; formate, acetate and their ammonium salts are in.

It also needs gas. Agilent’s 1260 ELSD takes 0.9 to 3.25 standard L/min, so a nitrogen supply is part of the installation. And the sample is destroyed, so the ELSD is always the last detector in line.

One thing it shows that UV does not: particles from the column. Agilent warns that an amino column with an aqueous mobile phase can shed packing and give a very high baseline.

The numbers

Steps (Agilent 1260 ELSD)
Nebuliser 25 to 90 °C, evaporator 25 to 120 °C, then a light source and a PMT 1
Light source (Agilent)
LED 480 nm (G4260B) or laser 405 nm (G4261B) 1
Gas (Agilent)
0.9 to 3.25 standard L/min, supplied at 4 to 6.7 bar 1
Eluent flow (Agilent)
0.2 to 5.0 mL/min 1
Mobile phase
Must be fully volatile (formate, acetate and their ammonium salts) 3
Sees (Agilent)
Any compound less volatile than the mobile phase, whatever its optical properties, so also compounds with no UV chromophore 2
Gradients (Agilent)
Used with gradients; response changes across the gradient, and a gas-flow timetable can compensate 2
Evaporator (Agilent)
80 to 90 °C for non-volatile compounds such as sugars; 20 to 30 °C for semi-volatile ones such as many drugs 2
Gas (Agilent)
Nitrogen above 98%; air only with non-flammable solvents 2
Position
Always the last detector in series 2
Sample
Destroyed in the detector 3

Same part, different names

Agilent
1260 Infinity II ELSD (G4260B), 1290 Infinity II ELSD (G7102A) 1

What goes wrong

  • High, noisy baseline after switching a UV method to ELSDCheck: A non-volatile buffer (such as phosphate) leaves solid residue after evaporation. ELSD needs a fully volatile mobile phase (Agilent). 3
  • Very high baseline with an amino columnCheck: Column shedding: the ELSD sees packing particles. Agilent names amino columns with aqueous mobile phase as prone to it. 2
  • No signal and the detector stopsCheck: Check the gas supply first. The Agilent ELSD has a gas shut-off valve and needs 4 to 6.7 bar inlet pressure. 1

Check yourself

Tap an answer to see if you are right and why.

What does an ELSD need from the mobile phase?

Sources

  1. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 58 (G4260B), pp. 92-93.
  2. Agilent Technologies, 2022, InfinityLab Evaporative Light Scattering Detectors User Manual (D0013647). Introduction, p. 8; operational parameters, p. 12; evaporator temperature, p. 48; real-time operation, p. 52; general considerations, pp. 55-56.
  3. Agilent Technologies, The LC Handbook. Detector types, p. 24; considerations for LC/MS and mobile phase modifiers, p. 74.