Sample manager

Takes a measured volume from the vial and puts it into the flowing mobile phase, the same way every time, without leaving any behind.

The sample manager has one job: the same volume, onto the column, every time, and nothing left behind for the next injection.

All of them work around a 6-port injection valve. While the sample is drawn, the pump flow goes straight past to the column. Then the valve turns, and the mobile phase pushes the sample out of the loop and onto the column.

Two things decide whether the numbers on your report are right: the precision of the volume (Agilent specifies below 0.25% RSD above 5 µL) and the carryover from the last sample (below 0.004% with needle wash). Both depend on parts in this module, so both are covered below.

The makers build it in two main ways, a fixed loop or a needle that is part of the loop. Start with that page, because it changes what you check when something goes wrong.

Point at a part in the drawing to open it.

The numbers

Also called
Autosampler, vialsampler, multisampler, sample manager, injector 1
Heart of it
A 2-position, 6-port injection valve 2
Injection range (Agilent 1260 Infinity III vialsampler)
0.1 to 100 µL in 0.1 µL steps 1
Injection precision (same)
Below 0.25% RSD of peak area from 5 to 100 µL; below 1% from 1 to 5 µL 1
Carryover spec (same)
Below 0.004% with needle wash (chlorhexidine) 1
Sample temperature
4 to 40 °C, with the optional sample thermostat (Agilent) or as standard (Waters ACQUITY) 1

Same part, different names

Agilent
Vialsampler, multisampler, autosampler (ALS) 1
Waters (ACQUITY)
Sample manager (SM-FTN, SM-FL) 3
Thermo Scientific (UltiMate 3000)
Autosampler (WPS-3000) 2
Shimadzu
Autosampler (SIL) 4
KNAUER
Autosampler (AZURA AS 6.1L) 5

What goes wrong

  • A peak in the blank after a strong sampleCheck: Carryover. Agilent lists six places it can hide: needle outside, needle inside, needle seat, sample loop, seat capillary and the injection valve. 1
  • Peaks smaller than expectedCheck: Thermo's list: wrong vial or too little sample, a vial filled to the top (a vacuum forms), a sticking syringe, a blocked needle, the wrong loop 6
  • Pressure high after the injectorCheck: Loosen the injector outlet fitting. If the pressure drops, the injector or its tubing is blocked. 6

Module check

8 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.

1.How many ports does a standard autosampler injection valve have?
2.In a flow-through needle design, what does the needle seat do?
3.In partial loop mode on a fixed loop, what is the most you should inject?
4.What sets the injection volume in most modern autosamplers?
5.What does the needle wash clean?
6.How full should a sample vial be, according to Thermo?
7.Click the part that switches the sample into the pump flow.
sample tray · 4 °C vial needle seat wash port sample loop metering device injection valve

Tap a part of the drawing. The labels are hidden.

8.Click the part that rinses the outside of the needle.
sample tray · 4 °C vial needle seat wash port sample loop metering device injection valve

Tap a part of the drawing. The labels are hidden.

Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.

Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →

Go deeper

Sources

  1. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Operating principle, pp. 15-18; Performance specifications, p. 27; Optimization for lowest carryover, p. 184.
  2. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Operating principle, PDF pp. 24-25; Injection valve, PDF p. 155.
  3. Waters, ACQUITY UPLC Sample Manager - Flow Through Needle: Operator's Overview and Maintenance Information (Rev C). Overview, p. 1.
  4. Shimadzu, SIL-30AC Autosampler Instruction Manual. Sample injection, p. 18.
  5. KNAUER, AZURA Autosampler AS 6.1L, Instructions (V6821). Injection principles, section 6.3.
  6. Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Smaller than expected peaks, PDF p. 19; pressure questions, PDF p. 7.