Sample manager

Injection designs: fixed loop and flow-through needle

Either the sample is pushed into a fixed loop, or the needle itself is part of the loop and stays in the flow. The design decides the volume range and where carryover hides.

There are two ways to get the sample into the flow.

Fixed loop. The needle draws sample and pushes it into a loop on the valve. The needle is never under high pressure. Waters’ ACQUITY sample manager in its default mode draws the injection volume plus 15 µL so that only the middle of the drawn plug goes into the loop.

Flow-through needle (Waters), split loop (Thermo), flow-through (Agilent). The needle is part of the loop. After drawing, it goes back into its seat, the valve turns, and the mobile phase flows through the needle itself. Thermo says this is loss-free: everything drawn is injected.

What it means at the bench:

  • In a flow-through design the needle inside and the loop are flushed by the mobile phase all through the run. Carryover then comes mostly from the outside of the needle and the needle seat.
  • In a flow-through design the loop and needle add to the dwell volume. Shimadzu’s SIL-30AC table gives its sample loop as 84.8 µL and its needle as 11.7 µL, both in the path from the mixer to the column. See Dwell volume in the Pump module.

The numbers

Flow-through needle (Waters SM-FTN)
The needle is part of the high-pressure sample flow path; extension loops add volume 1
Split loop (Thermo WPS-3000)
The needle is part of the sample loop; the drawn volume is injected completely 2
Flow-through (Agilent vialsampler)
Loop, needle inside, seat capillary and valve main path stay in the flow during the run 3
Fixed loop, partial loop with needle overfill (Waters ACQUITY default)
Draws the injection volume plus 15 µL (14 µL before, 1 µL after), then pushes the sample into the loop 4
Fixed loop (KNAUER AS 6.1L)
Three modes: full loop, partial loop, µL pickup 5

Same part, different names

Agilent
Flow-through design 3
Waters
Flow-through needle (FTN); fixed loop (FL) 1
Thermo Scientific
Split-loop injection 2

What goes wrong

  • Carryover that the needle wash does not fix (flow-through designs)Check: The needle seat. Agilent says to clean it with a solvent that dissolves the contamination, or replace it. 3
  • Gradient arrives later than on another systemCheck: In a flow-through design the loop and needle are part of the dwell volume. Thermo's bypass mode takes them out of the flow after the sample has passed. 2
  • Carryover went up after you switched on bypass (delay volume reduction)Check: Agilent: in bypass the sample path is not flushed by the gradient, so sample can stay in the loop, needle and seat. It does not recommend bypass for carryover-sensitive work. 7

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Waters, ACQUITY UPLC Sample Manager - Flow Through Needle: Operator's Overview and Maintenance Information (Rev C). Overview, p. 1.
  2. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Operating principle and bypass mode, PDF pp. 24-25 (manual pp. 16-17).
  3. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Injection sequence, p. 16; Optimization for lowest carryover, p. 184.
  4. Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). How sample flows, p. 1-7.
  5. KNAUER, AZURA Autosampler AS 6.1L, Instructions (V6821). Injection principles, section 6.3.
  6. Shimadzu, SIL-30AC Autosampler Instruction Manual. Holding capacity in the flow line, table, p. 18.