Sample manager

Needle wash

Rinses the outside of the needle after it leaves the vial, so the sample on it doesn't reach the seat or the next vial.

After the needle comes out of the vial, its outside is wet with sample. If that goes into the seat, or into the next vial, you get carryover. The needle wash rinses it off.

Agilent’s vialsampler dips the needle into a wash port that a peristaltic pump keeps full of fresh solvent. Its guidance on time: 2 to 3 seconds for routine work, 10 to 20 seconds when the analyte sticks to the needle. For the hardest cases it adds a wash vial with a stronger solvent, and that vial must be left uncapped, or the septum just wipes sample back onto the needle.

Waters’ ACQUITY uses two wash solvents, a weak one and a strong one, in a 3:1 ratio. Shimadzu can rinse the needle inside as well as outside, and then rinse the injection port.

Which solvent to use has its own page below, because that is where most carryover problems are solved.

The numbers

Agilent vialsampler wash port
Peristaltic pump at 5 mL/min; port volume about 0.5 mL, so it is replaced in about 6 s 1
Wash time (Agilent)
2 to 3 s for routine work; 10 to 20 s when the analyte sticks or carryover limits are low 1
Wash time (Agilent 1200 HiP)
At least 10 s, with a flush solvent that dissolves the sample 2
Waters ACQUITY
Two wash solvents: a weak wash and a strong wash, 3:1 by volume 3
Shimadzu SIL-30AC
External rinsing of the needle, internal rinsing, and rinsing of the injection port 4
Thermo WPS-3000
Wash port rinses the needle outside, then air from a membrane pump dries it 5

Same part, different names

Agilent
Automated needle wash, wash port, flush port 1
Waters
Weak wash and strong wash; inject/wash port 3
Shimadzu
Needle rinsing, rinsing port 4
Thermo Scientific
Wash port, wash liquid 5

What goes wrong

  • Carryover only after very concentrated samplesCheck: Lengthen the wash (Agilent: 10 to 20 s), or add a wash vial with a stronger solvent in the injector program 1
  • Carryover from a capped wash vialCheck: Agilent: leave the wash vial uncapped; sample left on the septum rides on the needle to the next vial 1

Check yourself

Tap an answer to see if you are right and why.

What does the needle wash clean?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Using the automated needle wash and Using the needle wash in the wash vial, pp. 184-185.
  2. Agilent Technologies, 1200 Series High Performance Autosampler User Manual (G1367-90011). General recommendations for lowest carry-over, p. 68.
  3. Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). Selecting weak wash and strong wash solvents, pp. 3-15 to 3-17.
  4. Shimadzu, SIL-30AC Autosampler Instruction Manual. Rinsing of the injection port, p. 17.