Sample manager
Needle wash
Rinses the outside of the needle after it leaves the vial, so the sample on it doesn't reach the seat or the next vial.
After the needle comes out of the vial, its outside is wet with sample. If that goes into the seat, or into the next vial, you get carryover. The needle wash rinses it off.
Agilent’s vialsampler dips the needle into a wash port that a peristaltic pump keeps full of fresh solvent. Its guidance on time: 2 to 3 seconds for routine work, 10 to 20 seconds when the analyte sticks to the needle. For the hardest cases it adds a wash vial with a stronger solvent, and that vial must be left uncapped, or the septum just wipes sample back onto the needle.
Waters’ ACQUITY uses two wash solvents, a weak one and a strong one, in a 3:1 ratio. Shimadzu can rinse the needle inside as well as outside, and then rinse the injection port.
Which solvent to use has its own page below, because that is where most carryover problems are solved.
The numbers
- Agilent vialsampler wash port
- Peristaltic pump at 5 mL/min; port volume about 0.5 mL, so it is replaced in about 6 s 1
- Wash time (Agilent)
- 2 to 3 s for routine work; 10 to 20 s when the analyte sticks or carryover limits are low 1
- Wash time (Agilent 1200 HiP)
- At least 10 s, with a flush solvent that dissolves the sample 2
- Waters ACQUITY
- Two wash solvents: a weak wash and a strong wash, 3:1 by volume 3
- Shimadzu SIL-30AC
- External rinsing of the needle, internal rinsing, and rinsing of the injection port 4
- Thermo WPS-3000
- Wash port rinses the needle outside, then air from a membrane pump dries it 5
Same part, different names
What goes wrong
- Carryover only after very concentrated samplesCheck: Lengthen the wash (Agilent: 10 to 20 s), or add a wash vial with a stronger solvent in the injector program 1
- Carryover from a capped wash vialCheck: Agilent: leave the wash vial uncapped; sample left on the septum rides on the needle to the next vial 1
Check yourself
Tap an answer to see if you are right and why.
Go deeper
- Carryover: where it hides and how to test itStudents Sample from one injection that turns up in the next. Test it with a blank after your strongest sample, then work through the places it can hide.
- Choosing the needle wash solventStudents The wash has to dissolve your analyte. Stronger is not always better: one Waters study found more carryover with 100% organic wash.
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Using the automated needle wash and Using the needle wash in the wash vial, pp. 184-185.
- Agilent Technologies, 1200 Series High Performance Autosampler User Manual (G1367-90011). General recommendations for lowest carry-over, p. 68.
- Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). Selecting weak wash and strong wash solvents, pp. 3-15 to 3-17.
- Shimadzu, SIL-30AC Autosampler Instruction Manual. Rinsing of the injection port, p. 17.
- Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Operating principle, PDF pp. 24-25.
New modules by email
All the modules are open now, and I keep adding parts and checks. Leave your email and I will send you what's new, including the Empower pages.
My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy