sample tray · 4 °C vial needle seat wash port sample loop metering device injection valve

Point at a part, then click to open it. Back to the whole system

Sample manager

Vials, caps, septa and plates

The container is part of the injection. Fill level, septum type and cap tightness all change what the needle draws.

The vial is the last thing you touch before the instrument takes over, and it can still spoil the injection.

Don’t fill it to the top. Thermo asks for about two-thirds liquid and one-third air. An over-full vial with a solid septum can form a vacuum as the needle draws, so less sample goes in and the peaks come out small. Thermo’s WPS-3000 can even move the tray a little while the needle is in the septum, to open the hole and let air in.

Septa and covers can block the needle. Shimadzu names septa as a cause of blocked needles, and Agilent warns against adhesive plate mats because the adhesive clogs the needle.

Thermo lists silicone, silicone/PTFE and slotted silicone/PTFE septa for its autosampler. A slotted septum lets air in easily.

The numbers

Fill level (Thermo)
About two-thirds liquid, one-third air; don't over-tighten caps 1
Why
An over-full vial can form a vacuum as the needle draws, so less sample is injected 1
Venting (Thermo WPS-3000)
The carousel can move slightly while the needle is in the septum, enlarging the hole to equalise pressure 2
Septa (Thermo)
Silicone, silicone/PTFE and slotted silicone/PTFE; other septa may block the capillaries 2
Plates (Waters ACQUITY)
ANSI/SBS footprint plates, up to 2.2 in high including covers 3
Plate covers (Agilent 1200 HiP)
Avoid adhesive closing mats: with no pre-punch needle, the adhesive clogs the needle 4

Same part, different names

Agilent
Vials, well plates, closing mats 4
Waters
Vials, microtiter plates, well covers 3
Thermo Scientific
Vials, well plates, septa 2

What goes wrong

  • Peaks smaller than expected from a full vialCheck: A vacuum in a full vial with an unslit septum. Fill to about two-thirds, or use slit septa or venting. 1
  • Needle blockedCheck: Pieces of septum or adhesive from plate mats. Shimadzu and Agilent both name these. 5

Check yourself

Tap an answer to see if you are right and why.

How full should a sample vial be, according to Thermo?

Sources

  1. Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Smaller than expected peaks, PDF p. 19.
  2. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Venting the vial, section 5.7.5; vials, caps and septa list and septa note, PDF pp. 37-38.
  3. Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). Sample manager, p. 1-7.
  4. Agilent Technologies, 1200 Series High Performance Autosampler User Manual (G1367-90011). Closing mats, p. 57.
  5. Shimadzu, HPLC Troubleshooting Guide. No peaks: blocked needle, p. 8.