Module
Detector
Turns what comes off the column into a signal. In most labs it is a UV detector: a lamp, a slit, a flow cell and a light sensor.
The detector is the last module the sample meets. Everything before it decides how good the separation is. The detector decides what you can see of it.
Agilent puts UV detectors in more than two-thirds of all LC systems, so most of this module is about UV. Inside a UV detector the light goes in one line: lamp, slit, flow cell, light sensor. The mobile phase leaves through a small restrictor to waste.
Three numbers describe any detector: how little it can see (sensitivity), how specific it is (selectivity), and over what range the signal stays proportional to concentration (linear range). For UV, Agilent gives 4 to 5 orders of magnitude of linear range.
When the compound has no UV absorbance, you need another type: fluorescence, refractive index or evaporative light scattering. Each has its own page here. The mass spectrometer gets its own section later.
Point at a part in the drawing to open it.
The numbers
- Most common type
- UV/Vis absorbance, used in more than two-thirds of LC systems 1
- Other types
- Fluorescence, refractive index, evaporative light scattering, mass spectrometry 1
- Keep the sample (nondestructive)
- UV, fluorescence, refractive index 1
- Destroy the sample
- Evaporative light scattering, mass spectrometry 1
- UV linear range
- 4 to 5 orders of magnitude 1
- Limit of detection
- Signal 3 times the noise 1
Same part, different names
What goes wrong
Module check
12 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.
Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.
Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →
Go deeper
- Lamp The light source. A deuterium lamp covers the UV; a diode array detector often adds a tungsten lamp for the visible range. Lamps get weaker with hours, and noise goes up.
- Slit and bandwidth The slit sets how wide a band of wavelengths reaches the sensor. Wider slit: more light, less noise. Narrower slit: sharper spectra.
- Flow cell The small window the mobile phase flows through while the light passes across it. Path length sets the signal; volume sets how much the peaks spread.
- VWD and diode array (DAD/PDA) A variable wavelength detector picks one wavelength before the cell. A diode array sends all the light through the cell, then splits it onto hundreds of diodes, so you get the full spectrum of every peak.
- Outlet restrictor (backpressure regulator) A small regulator or a length of narrow tubing on the detector outlet. It keeps a little pressure on the flow cell, so gas stays dissolved and does not make bubbles and spikes.
- Data rate and response time How many points per second the detector records, and how much it smooths them. Too slow and narrow peaks get flattened; too fast and the baseline gets noisy.
- Noise, drift and spikes What a bad baseline looks like tells you where to look: a slow drift, a regular wave, random fuzz or sharp spikes each point to a different part.
- Fluorescence detector (FLD) Shines one wavelength in and reads the light the compound gives back at a longer wavelength. Only fluorescent compounds (or derivatised ones) show up, so the baseline is clean.
- Refractive index detector (RI, RID) Sees almost anything, including sugars and polymers with no UV absorbance, by measuring how the mobile phase bends light. It is very sensitive to temperature, flow pulses and mobile phase changes.
- Evaporative light scattering detector (ELSD) Sprays the eluent into a gas stream, evaporates the solvent and measures the light scattered by the dry particles left. Sees compounds with no chromophore, works with gradients, but needs a fully volatile mobile phase.
Sources
- Agilent Technologies, The LC Handbook. Detector, the organ of perception, pp. 24-27; troubleshooting tables, pp. 127-128.
- Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Chapters 3 (UV detectors) and 4 (special detectors).