nalam.ca Learn
Sign in Join a group

Point at a part, then click to open it. Back to the whole system

Detector

Fluorescence detector (FLD)

Shines one wavelength in and reads the light the compound gives back at a longer wavelength. Only fluorescent compounds (or derivatised ones) show up, so the baseline is clean.

A UV detector measures the light a compound takes away. A fluorescence detector measures the light it gives back.

The detector sends one wavelength into the cell (excitation) and reads a longer wavelength coming out (emission). Agilent’s 1260 FLD sets both from 200 to 1200 nm and uses a xenon flash lamp, rated 4000 hours.

Only compounds that fluoresce give a signal, so it is a selective detector. Compounds that don’t fluoresce can be derivatised first; the Agilent handbook gives amino acid derivatisation in the autosampler as an example.

Two things to remember. Dissolved air quenches fluorescence, so degassing matters more here than with UV. And the cell is fragile: Agilent rates it at 20 bar.

Find your own wavelengths. Agilent’s manual shows the OPA derivative of alanine: the literature gives 340 nm excitation, a scan on the detector found the maximum at 220 to 240 nm.

The check for an FLD is the Raman band of water: Agilent uses it to test both sensitivity and wavelength accuracy.

The numbers

How it works (Agilent 1260 FLD)
One excitation and one emission wavelength, each set from 200 to 1200 nm 1
Lamp
Xenon flash lamp, 20 W normal or 5 W economy mode; 4000 h life (Agilent) 1
Flow cell (Agilent)
8 µL fused silica, 20 bar maximum 1
Sensitivity test (Agilent)
Raman band of water: excitation 350 nm, emission 397 nm 1
Wavelength accuracy (Agilent)
±3 nm, checked on the Raman band of water 1
Keeps the sample
Yes, nondestructive 3
Why emission is longer
Some absorbed energy is lost without light before the molecule emits 2
Keep apart
Emission at least 25 nm above excitation, or stray light shows in the spectra 2
Derivatives
Amino acids are measured as fluorescent OPA derivatives (Agilent's example); scan for the best wavelength 2
Dissolved air
Quenches fluorescence: degas 3

Same part, different names

Agilent
FLD (G7121A), FLD Spectra (G7121B) 1

What goes wrong

  • Response lower than expected and driftingCheck: Dissolved air quenches fluorescence (Agilent). Check the degasser and that the mobile phase is degassed. 3
  • Cell leaks after adding tubing or a second detector after itCheck: The Agilent FLD cell is rated only 20 bar. Keep the line after it short and open. 1

Check yourself

Tap an answer to see if you are right and why.

What lamp does Agilent's 1260 fluorescence detector use?

Sources

  1. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 52 (G7121A), pp. 82-84.
  2. Agilent Technologies, 2021, InfinityLab LC Series 1260 Infinity II Fluorescence Detectors User Manual (SD-29000222). How the detector operates, pp. 15-17; Fit Spectral Range, p. 65; finding the best wavelengths, pp. 116-117.
  3. Agilent Technologies, The LC Handbook. Autosampler programming, p. 23; detector types, p. 24; glossary, degassing, p. 147.

New modules by email

All the modules are open now, and I keep adding parts and checks. Leave your email and I will send you what's new, including the Empower pages.

My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy