Detector
Slit and bandwidth
The slit sets how wide a band of wavelengths reaches the sensor. Wider slit: more light, less noise. Narrower slit: sharper spectra.
The slit decides how much light gets through and how narrow a band of wavelengths it is.
It is a trade. A wider slit lets more light in, so the baseline is quieter. A narrower slit gives finer spectra, which you need when you compare peak spectra.
On a variable wavelength detector the slit is usually fixed (Agilent’s 1260 VWD: 6.5 nm). On a diode array you can often choose: Agilent’s DAD WR has 1, 2, 4, 8 or 16 nm. On a Waters PDA the setting is called resolution: 1.2 nm for spectra, 3.6 nm for a quieter baseline.
Write this setting into the method. Two labs running the same method at different bandwidths will not get the same baseline noise.
The numbers
- Agilent 1260 VWD (G7114A)
- Fixed slit, 6.5 nm typical 1
- Agilent 1260 DAD WR (G7115A)
- Programmable slit: 1, 2, 4, 8 or 16 nm 1
- Waters ACQUITY PDA
- 512 photodiodes, 1.2 nm optical resolution, 190 to 500 nm 2
- Less noise (Waters)
- 3.6 nm resolution gives a quieter baseline than 1.2 nm 2
- For spectral matching (Waters)
- Use 1.2 nm resolution 2
Same part, different names
What goes wrong
- Baseline noise too high for a low-level impurityCheck: If spectra are not needed, a wider resolution averages more diodes and lowers noise (Waters: 3.6 nm is quieter than 1.2 nm). 2
Check yourself
Tap an answer to see if you are right and why.
Sources
- Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 36 (VWD), p. 52; Table 38 (DAD WR), p. 55.
- Waters, ACQUITY UPLC PDA Detector Getting Started Guide (71500108703). Detector optics, p. 1-2; signal-to-noise and resolution, pp. 1-12 to 1-15; spectral contrast, p. 6-3.