Detector

Noise, drift and spikes

What a bad baseline looks like tells you where to look: a slow drift, a regular wave, random fuzz or sharp spikes each point to a different part.

The baseline is the detector telling you about the whole system. The shape of the problem is the clue.

  • Drift in one direction through a gradient: the mobile phase. A and B absorb differently at your wavelength.
  • A slow wave: temperature. A vent, or direct sunlight.
  • A regular ripple that keeps step with the pump: the pump. Air, check valves, degassing.
  • Constant fuzz: the lamp or a dirty flow cell.
  • Sharp spikes: bubbles in the cell, or electrical interference.

Thermo’s guide names the most common causes. For random noise: air in the system. For cyclic noise: a detector that was not given time to warm up and settle before the first injection.

So before you order a lamp, give the detector time, degas, and look at the shape again. The student page shows a simple test to tell the fluid path from the electronics.

The numbers

What noise depends on (Agilent)
Light intensity, pump pulsation, electronics, data rate, refractive index effects, temperature, flow rate, flow cell design 1
Limit of detection
Where the signal is 3 times the noise 1
Most common cause of random noise (Thermo)
Air in the system 2
Most common cause of cyclic noise (Thermo)
A detector not given enough time to stabilise 2
Agilent 1260 DAD WR drift
Below 0.9 × 10⁻³ AU/h at 254 nm 3

Same part, different names

Agilent
Short-term noise (ASTM), drift 3

What goes wrong

  • Baseline rises or falls through a gradientCheck: The two solvents absorb differently at your wavelength. Use solvents that absorb less, or a higher wavelength, or balance the additive between A and B (Agilent). 1
  • Slow wave over tens of minutesCheck: Room temperature. Use a column oven, and keep the detector out of air drafts (Agilent; for an RI detector, cover it). 1
  • Regular ripple in step with the pumpCheck: Pump pulses: purge air from the pump, clean or replace the check valves, degas (Agilent). 1
  • Constant fuzz that does not go awayCheck: Lamp or flow cell: replace the UV lamp, clean and flush the cell (Agilent). 1
  • Sharp spikesCheck: Bubbles (degas, add an outlet restrictor, tighten fittings) or electrical interference from equipment that cycles on and off (Agilent). 1

Check yourself

Tap an answer to see if you are right and why.

The baseline ripples in step with the pump strokes. Where do you look first?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, The LC Handbook. Detector sensitivity, p. 26; troubleshooting, baselines, pp. 127-128.
  2. Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Sections 3.3.1 to 3.3.3, baseline noise.
  3. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 38 (DAD WR), p. 55.