Detector
Noise, drift and spikes
What a bad baseline looks like tells you where to look: a slow drift, a regular wave, random fuzz or sharp spikes each point to a different part.
The baseline is the detector telling you about the whole system. The shape of the problem is the clue.
- Drift in one direction through a gradient: the mobile phase. A and B absorb differently at your wavelength.
- A slow wave: temperature. A vent, or direct sunlight.
- A regular ripple that keeps step with the pump: the pump. Air, check valves, degassing.
- Constant fuzz: the lamp or a dirty flow cell.
- Sharp spikes: bubbles in the cell, or electrical interference.
Thermo’s guide names the most common causes. For random noise: air in the system. For cyclic noise: a detector that was not given time to warm up and settle before the first injection.
So before you order a lamp, give the detector time, degas, and look at the shape again. The student page shows a simple test to tell the fluid path from the electronics.
The numbers
- What noise depends on (Agilent)
- Light intensity, pump pulsation, electronics, data rate, refractive index effects, temperature, flow rate, flow cell design 1
- Limit of detection
- Where the signal is 3 times the noise 1
- Most common cause of random noise (Thermo)
- Air in the system 2
- Most common cause of cyclic noise (Thermo)
- A detector not given enough time to stabilise 2
- Agilent 1260 DAD WR drift
- Below 0.9 × 10⁻³ AU/h at 254 nm 3
Same part, different names
- Agilent
- Short-term noise (ASTM), drift 3
What goes wrong
- Baseline rises or falls through a gradientCheck: The two solvents absorb differently at your wavelength. Use solvents that absorb less, or a higher wavelength, or balance the additive between A and B (Agilent). 1
- Slow wave over tens of minutesCheck: Room temperature. Use a column oven, and keep the detector out of air drafts (Agilent; for an RI detector, cover it). 1
- Regular ripple in step with the pumpCheck: Pump pulses: purge air from the pump, clean or replace the check valves, degas (Agilent). 1
- Constant fuzz that does not go awayCheck: Lamp or flow cell: replace the UV lamp, clean and flush the cell (Agilent). 1
- Sharp spikesCheck: Bubbles (degas, add an outlet restrictor, tighten fittings) or electrical interference from equipment that cycles on and off (Agilent). 1
Check yourself
Tap an answer to see if you are right and why.
Go deeper
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- Agilent Technologies, The LC Handbook. Detector sensitivity, p. 26; troubleshooting, baselines, pp. 127-128.
- Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Sections 3.3.1 to 3.3.3, baseline noise.
- Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 38 (DAD WR), p. 55.