Module

Data system (CDS)

The software that runs the instrument, records the detector signal, finds and integrates the peaks, calculates results and keeps the audit trail.

Every number you report comes out of this module. The detector gives a signal; the data system turns it into peaks, areas and results.

It does four jobs. It controls the instrument through a method. It records the detector signal. It processes the data: finds the peaks, integrates them, calibrates and calculates. And it keeps the record: who did what, when, and with which settings.

The last job is the one auditors look at. FDA calls a chromatographic record dynamic: change the baseline and the peaks get smaller or larger. So the audit trail has to show the integration parameters and every reprocessing, with the reason.

The four big packages are Empower, OpenLab CDS (and ChemStation), Chromeleon and LabSolutions. They do the same things with different words. The first page here puts the words side by side.

Point at a part of the screen in the drawing to open it.

The numbers

What integration does (Agilent)
Finds the start, apex and end of each peak, draws a baseline, and calculates area, height, width and symmetry 1
Audit trail (FDA)
For an HPLC run: user name, date and time, integration parameters, and any reprocessing with its reason 2
Why it matters (FDA)
A chromatographic record is dynamic: a changed baseline makes peaks smaller or larger 2

Same part, different names

Waters
Empower 3 3
Agilent
OpenLab CDS, OpenLab ChemStation 1
Thermo
Chromeleon 7 4
Shimadzu
LabSolutions 5

What goes wrong

  • Same raw data, different results from two analystsCheck: Different processing methods or manual integration. Compare the integration events and the audit trail of each result. 2

Module check

8 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.

1.In Empower, what is the list of injections to run called?
2.What should the peak width setting in the integrator be close to?
3.Two peaks touch. What does a drop line do?
4.A point-to-point calibration gives r = 1.000. What does that tell you about linearity?
5.Which width does the USP <621> plate number formula N = 5.54 (tR/Wh)² use?
6.For an HPLC run, what does FDA expect the audit trail to include?
7.Click the part of the screen where the baselines are drawn.
8.Click the part that records who changed what, and when.

Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.

Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →

Go deeper

Sources

  1. Agilent Technologies, 2022, OpenLab CDS Data Analysis Reference Guide, version 2.7 (D0013815). What is integration?, p. 13.
  2. US FDA, December 2018, Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry. Definitions: audit trail, p. 4; static and dynamic records, p. 5.
  3. Waters, 2017, Empower 3 Getting Started Guide (715005480). Empower concepts, p. 7.
  4. Thermo Fisher Scientific, 2017, Chromeleon 7.2 SR5 Quick Start Guide. Creating a sequence, p. 13.