The CDS turns peak area into an amount through a calibration curve. The fit type decides what it does outside the standards, and some fits look perfect by construction.

Integration gives you an area. Calibration turns that area into an amount, using standards of known amount.

Agilent’s reference guide says it plainly: the response is rarely in direct proportion to the amount, so you calibrate. With an external standard you compare sample area with standard area. With an internal standard you compare the ratio of the analyte peak to an internal standard peak added to every solution.

The fit type matters more than people think. Two lines from the Empower theory guide:

  • A linear fit extrapolates. A sample above your top standard still gets a number.
  • A point-to-point or cubic spline fit passes through every standard, so r = 1 and the standard error is 0 by construction. A perfect r there tells you nothing.

So read the fit type before you read the r value.

The numbers

Why calibrate (Agilent)
Response is rarely in direct proportion to amount; reference standards are needed 1
Types (Agilent)
External standard (ESTD), internal standard (ISTD), response factor from another compound, fixed factor 1
Multi-level (Agilent)
Several standards when the detector response is not linear 1
Response factor (Agilent)
Response / amount by default; can be switched to amount / response 1
Linear fits (Empower)
Extrapolate above the highest and below the lowest standard 2
Non-linear fits (Empower)
Report only from 0 to the highest standard 2
Point-to-point (Empower)
Passes through every point, so r = 1 and standard error = 0 by construction 2
Quadratic (Empower)
At least 3 points; 5 strongly recommended 2

Same part, different names

Agilent
Calibration table, ESTD/ISTD, RF 1
Waters
Calibration curve, fit type 2

What goes wrong

  • Sample result above the top standard, reported without a flagCheck: Linear fits extrapolate (Empower). Check the method's range and flag or dilute samples above the highest standard. 2
  • Calibration shows r = 1.000 every timeCheck: Point-to-point or cubic spline: they pass through every point by construction, so r says nothing about linearity (Empower). 2

Check yourself

Tap an answer to see if you are right and why.

A point-to-point calibration gives r = 1.000. What does that tell you about linearity?

Sources

  1. Agilent Technologies, 2022, OpenLab CDS Data Analysis Reference Guide, version 2.7 (D0013815). What is calibration?, p. 99; calibration curve and response factor, pp. 100-101.
  2. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Calibration curve fit types, pp. 88-94.