Data system

Finding peaks: peak width and threshold

Two settings decide what the software calls a peak: the expected peak width, and how big a rise above the noise counts. Set them wrong and peaks vanish or noise becomes peaks.

Before the software can integrate a peak, it has to decide that it is a peak. Every CDS uses two settings for that.

Peak width. The software looks for rises of about the width you tell it. Agilent: set it close to the real peak width. Shimadzu: peaks narrower than Width are treated as noise. So set it from your narrowest real peak.

Threshold, or slope. How steep a rise has to be before it counts. Too low and baseline noise becomes peaks; too high and small impurities disappear. Shimadzu has a slope test that measures the baseline for you, but warns not to use it for gradients, because it doesn’t see the drift that comes later in the run.

Peaks get wider later in an isocratic run. Agilent’s integrator updates the width as it goes; Shimadzu uses T.DBL to double the width at set times.

Change these settings in the processing method, not peak by peak. The page on manual integration says why.

The numbers

Agilent peak width
Must be close to the real peak width; it separates peaks from noise 1
Agilent minimum
A peak needs at least 10 data points 1
Agilent: other settings
Slope sensitivity, area reject, height reject, shoulder detection 1
Shimadzu Width
Half-height width; anything narrower is treated as noise 2
Shimadzu Slope
Set from a slope test on the baseline; for gradients, set it by hand above the drift 2
Shimadzu T.DBL
Doubles Width and halves Slope sensitivity at set times, for peaks that broaden 2
Empower
Peak Width and Detection Threshold; blank = automatic values 3
Empower threshold
Based on baseline peak-to-peak noise; if too many peaks, raise it in steps of 10 3

Same part, different names

Agilent
Peak width, slope sensitivity 1
Shimadzu
Width, Slope, Drift, T.DBL, Min. Area 2
Waters
Peak Width, Detection Threshold (ApexTrack) 3

What goes wrong

  • Baseline bumps reported as peaks in a gradient runCheck: Threshold or slope too sensitive. Shimadzu: the slope test does not see gradient drift, so set Slope above the drift by hand. 2
  • Small impurity next to the main peak not foundCheck: Peak width set too wide, or the threshold too high. Agilent: peak width must match the real peaks. 1
  • Empower's automatic peak width looks wrongCheck: Empower: it is too wide when the largest peak is saturated or co-eluting, too narrow when it is noisy. Exclude the void and re-equilibration. 3

Check yourself

Tap an answer to see if you are right and why.

What should the peak width setting in the integrator be close to?

Sources

  1. Agilent Technologies, 2022, OpenLab CDS Data Analysis Reference Guide, version 2.7 (D0013815). Integrator capabilities, p. 14; peak width and recognition filters, pp. 21-23.
  2. Shimadzu, 2017, LabSolutions Data Acquisition & Processing Theory Guide, version 5.93 (223-60090G). Peak integration parameters, pp. 1-4; slope test, p. 6.
  3. Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). ApexTrack peak width and detection threshold, pp. 13, 20-26.