Data system
System suitability in the CDS
The CDS calculates plates, resolution, tailing, S/N and %RSD for you. Which formula it uses is a setting, and the pharmacopeia makes it your job to check.
System suitability is the check that the system works today, before you trust today’s results. The CDS does the arithmetic: plates, resolution, symmetry (tailing), signal-to-noise, %RSD of replicate injections.
But each number has more than one formula. USP <621> calculates plates and resolution from the width at half height. Empower’s “USP plate count” uses the width at the tangent. Agilent offers four ways to measure noise. Same chromatogram, different numbers.
USP puts the responsibility on you: the user must make sure the software’s calculation is the one the chapter requires.
So when a method moves between labs or between CDS, check the formula behind each SST field, not just the limit. The student page shows how.
The numbers
- Plate number (USP)
- N = 5.54 (tR / Wh)², from the width at half height 1
- Resolution (USP)
- Rs = 1.18 (tR2 − tR1) / (Wh1 + Wh2) 1
- Symmetry (tailing) factor (USP)
- As = W0.05 / 2d, at 5% of peak height 1
- USP on software
- The user must make sure the software's calculations match the USP requirements 1
- Agilent OpenLab
- Reports k', symmetry, width, plates, resolution and S/N per calibrated peak; USP, EP or JP definitions 2
- Agilent noise options
- 6 × SD of the drift line, peak-to-peak, ASTM E685, or RMS 2
- Empower
- Width at tangent is used for USP resolution and USP plate count 3
Same part, different names
What goes wrong
- Plate count differs between two CDS for the same chromatogramCheck: Different width: half-height (USP formula) or tangent (Empower's USP plate count). Check which field the method reports. 3
- S/N passes in one CDS, fails in anotherCheck: Noise is calculated differently: Agilent offers four noise methods. Match the method to the pharmacopeia definition. 2
Check yourself
Tap an answer to see if you are right and why.
Go deeper
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- United States Pharmacopeia, USP General Chapter <621> Chromatography (harmonized text, official 1 Dec 2022). System suitability: definitions of plate number, resolution, symmetry factor; note on software calculations.
- Agilent Technologies, 2022, OpenLab CDS Data Analysis Reference Guide, version 2.7 (D0013815). System suitability, pp. 173-175; formulas and Table 9, pp. 188-190.
- Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Width at tangent, p. 21.