Data system
Audit trail and data integrity
A computer-made, time-stamped record of who created, changed or deleted data, when, and why. In a regulated lab it is reviewed with the results, like cross-outs on paper.
The audit trail is the part of the data system that watches you. Every creation, change and deletion of a record gets a line: who, when, what, and why.
FDA’s definition is short: a secure, computer-generated, time-stamped record that lets someone reconstruct what happened to a result. For an HPLC run, FDA lists the user, the date and time, the integration parameters, and any reprocessing with its reason.
EU Annex 11 says the same for Europe, and adds that audit trails must be readable and regularly reviewed.
Each CDS builds it in its own way. Thermo’s Chromeleon keeps an audit trail with every injection and stores every version of a method or result, so two versions can be compared. Empower keeps audit trail and sign-off information in the project.
The audit trail only protects you if someone reads it. The student page covers who reviews it, how often, and what to look for.
The numbers
- Definition (FDA)
- Secure, computer-generated, time-stamped record to reconstruct the creation, change or deletion of an electronic record 1
- For an HPLC run (FDA)
- User name, date and time of the run, integration parameters, reprocessing details and the reason 1
- Annex 11
- GMP-relevant changes and deletions recorded, with the reason; audit trails readable and regularly reviewed 2
- Electronic signature (Annex 11)
- Same weight as a handwritten one, permanently linked to the record, with date and time 2
- Chromeleon (Thermo)
- Audit trail per injection, versions of every data object, comparison between versions, logon and failed logon tracked 3
- Empower
- A project can hold audit trail and sign-off information with its methods and results 4
Same part, different names
What goes wrong
- Results signed off, audit trail never openedCheck: FDA: whoever reviews the record reviews the audit trail of the changes to it, at the same time. 1
- Test injections of a sample before the official sequenceCheck: FDA calls using a real sample in test, prep or equilibration runs to get a result a violative practice (testing into compliance). 1
Check yourself
Tap an answer to see if you are right and why.
Go deeper
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- US FDA, December 2018, Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry. Definitions: audit trail, p. 4; questions 7 and 8, p. 8; question 13, p. 11.
- European Commission, 2011, EudraLex Volume 4, Annex 11: Computerised Systems. Sections 9 (audit trails) and 14 (electronic signature).
- Thermo Fisher Scientific, 2018, Data Integrity: Technical Controls to Demonstrate Trust, Using Chromeleon 7 CDS (White Paper 80077). Audit trail review and versioning, pp. 5-7.
- Waters, 2017, Empower 3 Getting Started Guide (715005480). Empower concepts: project, p. 7.
New modules by email
All the modules are open now, and I keep adding parts and checks. Leave your email and I will send you what's new, including the Empower pages.
My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy