Module

Column manager

Holds the column at a steady temperature, and on many systems can switch between columns. The column is where the separation happens.

The column manager does two jobs: keep the column at one temperature, and on many systems, switch between columns.

Temperature matters more than it looks. Dolan’s troubleshooting guide puts it at 1 to 3% change in retention time for every degree. A lab can feel steady at the thermostat on the wall while the air from a vent blows straight onto the HPLC. A column oven takes that out of the picture.

Agilent calls the column thermostat an often underrated component, and it is the part that makes a method give the same retention times on Monday morning and Friday afternoon, and in another lab.

Inside this module you find the oven itself, the preheater that brings the mobile phase to temperature before the column, the guard column, the column, the connections, the switching valves and, on some columns, a memory chip.

Point at a part in the drawing to open it.

The numbers

Also called
Column manager, column compartment, column oven, column thermostat (TCC) 1
Temperature range (Waters ACQUITY column manager)
4 to 90 °C, set in 0.1 °C steps; no lower than 25 °C below room temperature 2
Typical column oven
Heats the column surroundings, usually up to 80 °C 1
Why it matters
Retention time changes 1 to 3% for every 1 °C change in column temperature 3
Columns held (Waters ACQUITY)
Two columns of 50, 100 or 150 mm, or four 50 mm columns; up to 4.6 mm ID 2

Same part, different names

Agilent
Thermostatted column compartment (TCC), multicolumn thermostat 1
Waters
Column manager, column heater 2

What goes wrong

  • Retention times drift through the dayCheck: No column oven, or the instrument sits under an air vent. Dolan: 1 to 3% per °C; use a column oven and move the system away from vents. 3
  • Rising pressure, drifting retention and more tailing over weeksCheck: Column ageing. Swap in a new column to confirm. Dolan treats 500 to 2000 injections as a satisfactory column life. 3

Module check

7 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.

1.By about how much can retention time change for each 1 °C change in column temperature?
2.What does a preheater do?
3.If you halve the particle size, what happens to the backpressure (same flow, same column size)?
4.Where does a guard column go?
5.Most silica-based columns are stable in which pH range?
6.Click the part that holds the column at a set temperature.

Tap a part of the drawing. The labels are hidden.

7.Click the short cartridge that protects the column from particles and strongly held sample.

Tap a part of the drawing. The labels are hidden.

Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.

Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →

Go deeper

Sources

  1. Agilent Technologies, The LC Handbook. The column thermostat, pp. 23-24; column compartment, p. 24.
  2. Waters, ACQUITY UPLC Column Manager: Operator's Overview and Maintenance Information. Features, active preheater and major components, pp. 1-1 to 1-4.
  3. ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Column temperature changes, section 2.3; column chemistry changes, section 2.2.