The nuts and ferrules at each end of the column. A wrong stem depth leaves a small void that broadens peaks; overtightening damages the column.
The column is only as good as its two connections. Two things go wrong.
A void. Column ports are not all the same depth. A ferrule that was swaged in one column’s port sits a little short in another, and the gap becomes a tiny mixing chamber. Optimize shows this for fixed-stem fittings and recommends a floating or spring-loaded stem that finds the bottom of any port. The result of a void is broader, tailing peaks on everything, which looks like a column problem but isn’t.
Overtightening. Agilent’s column care list asks for a short wrench on the 3/8 in end nuts and no more force than needed.
When you swap columns, remake the connections rather than assuming the old ferrule fits. The Tubing and fittings pages go further into nuts, ferrules and stem depth.
The numbers
- End nuts
- Columns have 3/8 in end nuts; use a short-handled wrench and don't overtighten (Agilent) 1
- Fixed-stem fittings (Optimize)
- A pre-swaged or one-piece fitting with a fixed stem can sit at the wrong depth and add dead volume 2
- Floating stem (Optimize)
- A spring-loaded or floating stem seats at zero dead volume in any port 2
- Agilent quick-connect
- A-line Quick Connect fittings, to avoid small leaks at column connections 1
- Waters ACQUITY column manager
- Plumb with the preconfigured tubing lengths, to reduce dead volume 3
What goes wrong
- Every peak broader or tailing after a column changeCheck: A void at a fitting: the ferrule set for a different column's port depth. Remake the connection, ideally with a floating-stem fitting. 2
- Small leak at the column inlet under pressureCheck: Tighten gently. Agilent warns against overtightening the 3/8 in end nuts. 1
Sources
- Agilent Technologies, The LC Handbook. Column care tips (end fittings), p. 119; column compartment, p. 24.
- Optimize Technologies, Guard Columns: guide to Optimize guard columns (OPTI-GUARD, OPTI-LYNX, EXP). Extra column volume, Figures 1 and 2, p. 1.
- Waters, ACQUITY UPLC Column Manager: Operator's Overview and Maintenance Information. Configurations, p. 1-4.