Column manager

Guard column and inline filter

A short cartridge or filter in front of the column that catches particles and strongly held sample material, so the expensive column lasts longer.

A guard column is a short column, usually packed with the same material, placed just before the analytical column. It catches particles and the sample components that would otherwise stick to the top of the column. Agilent calls it an economic decision: a guard costs little, a column a lot.

Agilent’s rule of thumb for when to change it: when the plate number, pressure or resolution change by more than 10%, and better sooner than later.

An inline filter is the smaller cousin: just a frit in a holder. Agilent matches the frit to the particles, 2 µm for a 3.5 µm column and 0.5 µm for 1.8 µm. On very small UHPLC columns, a guard can cost efficiency, so Agilent suggests a low-volume filter instead.

Either way, connect it well. Optimize points out that a fixed tube stem set at the wrong depth leaves a small void, and that void shows up as worse peaks.

The numbers

Where
Right before the analytical column (Agilent); between the injection valve and the column (Optimize) 1
Choose
Same packing as the column, and a similar inner diameter, so the separation does not change 1
When to replace (Agilent)
When plate number, pressure or resolution change by more than 10% 1
Inline filter frit (Agilent)
2 µm for a 3.5 µm column; 0.5 µm for a 1.8 µm column 1
Inline filter (Dolan)
A 0.5 µm inline filter; replacing its frit is quick and cheap 2
Very small columns (Agilent)
A guard can cost efficiency on very low-volume columns; use a low-volume inline filter instead 1
Guard sizes (Optimize OPTI-GUARD)
1.0, 2.1, 3.0 and 4.6 mm ID × 5 mm; up to 400 bar 3

Same part, different names

Waters
VanGuard pre-column, VanGuard FIT cartridge 4
Optimize Technologies
OPTI-GUARD, OPTI-LYNX, EXP guard 3

What goes wrong

  • Pressure rising slowly over daysCheck: A blocked inline filter frit or guard. Replace the cheap part first (Dolan) 2
  • Peak shape worse after fitting a guardCheck: Dead volume at the guard-to-column joint, for example a fixed tube stem at the wrong depth (Optimize) 3

Check yourself

Tap an answer to see if you are right and why.

Where does a guard column go?

Sources

  1. Agilent Technologies, The LC Handbook. Inline filters, p. 117; Guard columns, p. 118.
  2. ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Blocked frits and inline filters, p. 9; troubleshooting table, p. 5.
  3. Optimize Technologies, Guard Columns: guide to Optimize guard columns (OPTI-GUARD, OPTI-LYNX, EXP). Why use a guard column; extra column volume; guide to Optimize guard columns, pp. 1-2.
  4. Waters, ACQUITY UPLC and ACQUITY Premier Columns: Care and Use Manual (720008980). VanGuard and VanGuard FIT column protection.