Column manager
The column
A steel tube packed with small particles. Length, inner diameter and particle size decide efficiency and pressure; the bonded phase decides selectivity.
The column is where the separation happens. Everything else in the HPLC is there to deliver a sample and a steady mobile phase to it, and to read what comes out.
Four numbers on the label tell you most of what you need: length, inner diameter, particle size and phase.
- Particle size sets efficiency and pressure together. Agilent’s rule: halve the particle size and the plates double, but the pressure goes up four times. That is why sub-2 µm columns need a UHPLC.
- Length adds plates and time in equal measure, but resolution only by about 1.4 × when you double it.
- The phase sets selectivity. Most methods use C18.
Most silica columns are happy from pH 2 to 8. Below 2 the bonded phase hydrolyses; above 8 the silica dissolves. Dolan’s example of what that looks like: a non-endcapped column at pH 9 that suddenly gave fronting peaks after about 500 injections.
The deeper pages cover pressure and particle size, and care and storage.
The numbers
- Most used phase
- C18 (octadecylsilane, ODS); Agilent cites surveys where 95% of chromatographers use reversed phase 1
- HPLC particles
- 3 to 5 µm 1
- UHPLC particles
- Below 3 µm, run above 400 bar (6000 psi) 1
- Halve the particle size
- Plate number doubles, pressure goes up 4 times (about 1/dp²) 1
- Double the length
- Plates and time double, but resolution improves only about 1.4 times 1
- Pore size for small molecules
- Small pores, 60 to 120 Å 1
- Silica columns, pH
- Stable from pH 2 to 8 for most (Dolan); Agilent advises pH 2 to 7 for longest life 2
- Life
- 500 to 2000 injections is satisfactory (Dolan) 2
Same part, different names
What goes wrong
- Peak fronting appears suddenly after a few hundred injectionsCheck: A void or channel in the bed. Dolan's example: a non-endcapped column run at pH 9 dissolved its silica after about 500 injections. 2
- Pressure much higher after moving to a smaller-particle columnCheck: Expected: about 1/dp², so 5 µm to 2.5 µm is about 4 times the pressure at the same flow 1
Check yourself
Tap an answer to see if you are right and why.
Go deeper
- Pressure, particle size and column sizeStudents Five things set the pressure: viscosity, flow, length, column radius and particle size. Particle size counts most, as one over its square.
- Column care, flushing and storageStudents Flush out buffer, store in organic solvent, stay inside the pH and pressure limits, and a column lasts its 500 to 2000 injections.
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- Agilent Technologies, The LC Handbook. Resolution, p. 9; pressure equation, p. 10; reversed phase, p. 29; UHPLC columns, p. 32; column packings, pp. 35-36; Column care tips, p. 119.
- ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Fronting peaks, section 1.5; column chemistry changes, section 2.2.
- Waters, ACQUITY UPLC and ACQUITY Premier Columns: Care and Use Manual (720008980). Column care and use manual (720008980), introduction.