Column manager

The column

A steel tube packed with small particles. Length, inner diameter and particle size decide efficiency and pressure; the bonded phase decides selectivity.

The column is where the separation happens. Everything else in the HPLC is there to deliver a sample and a steady mobile phase to it, and to read what comes out.

Four numbers on the label tell you most of what you need: length, inner diameter, particle size and phase.

  • Particle size sets efficiency and pressure together. Agilent’s rule: halve the particle size and the plates double, but the pressure goes up four times. That is why sub-2 µm columns need a UHPLC.
  • Length adds plates and time in equal measure, but resolution only by about 1.4 × when you double it.
  • The phase sets selectivity. Most methods use C18.

Most silica columns are happy from pH 2 to 8. Below 2 the bonded phase hydrolyses; above 8 the silica dissolves. Dolan’s example of what that looks like: a non-endcapped column at pH 9 that suddenly gave fronting peaks after about 500 injections.

The deeper pages cover pressure and particle size, and care and storage.

The numbers

Most used phase
C18 (octadecylsilane, ODS); Agilent cites surveys where 95% of chromatographers use reversed phase 1
HPLC particles
3 to 5 µm 1
UHPLC particles
Below 3 µm, run above 400 bar (6000 psi) 1
Halve the particle size
Plate number doubles, pressure goes up 4 times (about 1/dp²) 1
Double the length
Plates and time double, but resolution improves only about 1.4 times 1
Pore size for small molecules
Small pores, 60 to 120 Å 1
Silica columns, pH
Stable from pH 2 to 8 for most (Dolan); Agilent advises pH 2 to 7 for longest life 2
Life
500 to 2000 injections is satisfactory (Dolan) 2

Same part, different names

Agilent
ZORBAX, Poroshell, InfinityLab columns 1
Waters
ACQUITY UPLC columns (BEH, CSH, HSS) 3

What goes wrong

  • Peak fronting appears suddenly after a few hundred injectionsCheck: A void or channel in the bed. Dolan's example: a non-endcapped column run at pH 9 dissolved its silica after about 500 injections. 2
  • Pressure much higher after moving to a smaller-particle columnCheck: Expected: about 1/dp², so 5 µm to 2.5 µm is about 4 times the pressure at the same flow 1

Check yourself

Tap an answer to see if you are right and why.

1.If you halve the particle size, what happens to the backpressure (same flow, same column size)?
2.Most silica-based columns are stable in which pH range?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, The LC Handbook. Resolution, p. 9; pressure equation, p. 10; reversed phase, p. 29; UHPLC columns, p. 32; column packings, pp. 35-36; Column care tips, p. 119.
  2. ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Fronting peaks, section 1.5; column chemistry changes, section 2.2.
  3. Waters, ACQUITY UPLC and ACQUITY Premier Columns: Care and Use Manual (720008980). Column care and use manual (720008980), introduction.