Column manager
Column switching valve
Valves at the inlet and outlet of the compartment that send the flow to one of several columns, or around them. Used for method scouting and for keeping two methods ready.
A column switching valve lets one HPLC hold more than one column and pick between them from the method.
On Waters’ ACQUITY column manager there are two of them, one at the inlet and one at the outlet, so the flow can go through column 1, column 2, a bypass around all columns, or to waste. Agilent lists the same kinds of uses: automated method development, column switching, enrichment, and regenerating one column while another runs.
In a QC lab the everyday use is simpler: two methods on two columns, both installed, so nobody has to swap columns between sequences.
The numbers
- Waters ACQUITY column manager
- Inlet and outlet switching valves give programmable access to any installed column, a bypass, and waste 1
- Columns (same)
- Two columns (50, 100 or 150 mm) or four 50 mm columns 1
- Guards (same)
- Guard columns or filters only in the two-column set-up 1
- Agilent
- Compartments can house several columns and valves for method development, column switching, enrichment or alternating regeneration 2
- Wear part (Waters)
- Switching-valve cartridge, replaceable 1
What goes wrong
Sources
- Waters, ACQUITY UPLC Column Manager: Operator's Overview and Maintenance Information. Features, p. 1-1; major components and configurations, pp. 1-3 to 1-4; replacing the switching-valve cartridge, chapter 4.
- Agilent Technologies, The LC Handbook. Column compartment, p. 24.