Column manager

Preheater

Warms the mobile phase to column temperature before it reaches the column, so the column is the same temperature from end to end.

A column oven heats the column. But the mobile phase arrives from the sample manager at room temperature, and at higher flow rates it can reach the column before it has warmed up. Then the inlet of the column is cooler than the rest.

The preheater fixes that by warming the solvent first. Waters’ ACQUITY uses an active preheater, an electric heater right on the column inlet tubing, so the separation runs at the column temperature, not at the temperature of the incoming solvent. Agilent’s compartments use solvent heat exchangers for the same job.

The trade-off is volume. Anything between the injector and the column adds to the extra-column volume, so these parts are built small.

The numbers

Waters active preheater
An electric heater on the column inlet tubing 1
Why (Waters)
High flow separations then happen at the column temperature, not at the temperature of the incoming solvent 1
Design (Waters)
Fast heating and low volume, to keep gradient delay and extra-column band spreading small 1
Agilent
Modern column compartments have solvent heat exchangers that heat the solvent to the set temperature 2

What goes wrong

  • A method moved after you switched from a column stabilizer to the active preheater (Waters)Check: Waters says the column temperature may need adjusting when you run a method developed with a column stabilizer on the active preheater 1
  • Peaks broader after a heat exchanger with long, wide tubing was addedCheck: The heat exchanger is part of the extra-column volume. Agilent sells heat exchanger capillaries in 0.12 and 0.075 mm ID; the narrower one adds less volume. 2

Check yourself

Tap an answer to see if you are right and why.

What does a preheater do?

Sources

  1. Waters, ACQUITY UPLC Column Manager: Operator's Overview and Maintenance Information. Active preheater, p. 1-2; major components, p. 1-3.
  2. Agilent Technologies, The LC Handbook. The column thermostat, p. 23; heat exchanger capillaries in the parts list, p. 184.

New modules by email

All the modules are open now, and I keep adding parts and checks. Leave your email and I will send you what's new, including the Empower pages.

My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy