Column manager

Column oven and temperature

Holds the column at a set temperature. Simple ovens heat the air or a block; Peltier thermostats can heat and cool.

The oven keeps the column at one temperature so the retention times stay put. Dolan’s figure is 1 to 3% per degree, so a 2 °C swing in the lab can move a 10-minute peak by about 12 to 36 seconds.

There are two kinds. A column oven or heater warms the space around the column, usually up to 80 °C. A Peltier thermostat can also cool, and changes temperature quickly, which is why it is used for method development. Waters’ ACQUITY column manager runs 4 to 90 °C, but cooling stops at 25 °C below room temperature.

Two habits from Agilent’s column care list: stay below the column’s maximum temperature (most reversed-phase columns take about 60 °C), and when you run hot, heat and cool with the flow running.

The numbers

Simple oven or heater
Heats the column surroundings, usually up to 80 °C 1
Peltier thermostat
Heats and cools, changes temperature faster; used in method development 1
Waters ACQUITY column manager
4 to 90 °C, each trough set separately in 0.1 °C steps 2
Cooling limit (same)
No lower than 25 °C below room temperature 2
Most reversed-phase columns
Withstand up to about 60 °C; check the column's own limit 1
Retention vs temperature
1 to 3% per 1 °C 3

What goes wrong

  • Retention times move during the day, the column runs at ambientCheck: Use the column oven, and move the system away from heating or air-conditioning vents (Dolan) 3
  • A method runs at a different retention on another systemCheck: Agilent: temperature control is one of the things that decide whether a method transfers. Check the setpoint and whether the mobile phase is preheated. 1
  • Column damaged after running hotCheck: Raise the temperature gradually, only with flow running, and cool down with flow on (Agilent) 1

Check yourself

Tap an answer to see if you are right and why.

By about how much can retention time change for each 1 °C change in column temperature?

Sources

  1. Agilent Technologies, The LC Handbook. The column thermostat, pp. 23-24; temperature limits, p. 81; Column care tips table, p. 119; method transfer, p. 192.
  2. Waters, ACQUITY UPLC Column Manager: Operator's Overview and Maintenance Information. Features, p. 1-1.
  3. ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Column temperature changes, section 2.3.