Point at a part, then click to open it. Back to the whole system

Module

Pump

Pulls solvent from the bottles and pushes it through the column at a set flow rate, against hundreds of bar.

The pump does one job: a steady flow, whatever the pressure. The column pushes back with hundreds of bar, and the pump has to deliver the same flow rate through all of it.

Most analytical HPLC pumps are dual-piston pumps. In the Agilent design, which the drawing follows, solvent goes in through the inlet valve into chamber 1, out through the outlet valve, through a damper, into chamber 2, and out past the purge valve to the rest of the system. The two pistons are driven in opposite directions, and the first moves at twice the speed of the second.

Every stroke gives a small pressure pulse. Agilent’s 1260 specification allows below 2% or below 3 bar. When the pulse grows past that, the cause is usually in this module: a check valve, a seal, or air.

Point at a part in the drawing to open it.

The numbers

Design (Agilent 1260)
Dual piston in series: two chambers, one sapphire piston each 1
Stroke volume, first piston
20 to 100 µL, set by the flow rate 1
Flow range (Agilent 1260)
1 µL/min to 10 mL/min 1
Pressure (Agilent 1260 quaternary)
Up to 600 bar 1
Pressure pulsation spec
Below 2% amplitude, or below 3 bar 1

What goes wrong

  • Pressure goes up and downCheck: Shimadzu's list: air bubbles, worn seals, check valves, leaks, poor degassing. A gradient changing viscosity is normal. 2
  • Pump will not primeCheck: ASI's list: back pressure in the head (open the purge valve), a check valve fitted upside down, a clogged inlet filter or outlet valve filter 3
  • Erratic pressure, and no leak you can seeCheck: Air leaking in at the inlet line. ASI: the intake side is under slight vacuum, so air leaks in and nothing drips out. 3

Module check

10 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.

1.Where are the solvents mixed in a quaternary (low-pressure mixing) pump?
2.What is the piston in an Agilent 1260 pump head made of?
3.What does a check valve in the pump head do?
4.Agilent's 1260 pump uses black PTFE piston seals as standard. When does Agilent call for yellow PE seals?
5.When does Agilent strongly recommend the seal wash function?
6.What is the purge valve for?
7.Where does the inline filter sit?
8.What pressure pulsation does Agilent specify for the 1260 pump?
9.Click a one-way valve: it lets solvent into the chamber and stops it flowing back.

Tap a part of the drawing. The labels are hidden.

10.Click the valve you open to prime the pump, so the flow goes to waste and not through the column.

Tap a part of the drawing. The labels are hidden.

Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.

Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →

Go deeper

Sources

  1. Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, pp. 19-22; How does variable stroke volume work?, p. 26; Specifications, PDF p. 42.
  2. Shimadzu, HPLC Troubleshooting Guide. Fluctuating pressure, p. 21.
  3. Analytical Scientific Instruments (ASI), Trouble Shooting Guide: pump replacement parts (white page). Erratic pressure and Failure to prime, pp. 1-3.