Module
Pump
Pulls solvent from the bottles and pushes it through the column at a set flow rate, against hundreds of bar.
The pump does one job: a steady flow, whatever the pressure. The column pushes back with hundreds of bar, and the pump has to deliver the same flow rate through all of it.
Most analytical HPLC pumps are dual-piston pumps. In the Agilent design, which the drawing follows, solvent goes in through the inlet valve into chamber 1, out through the outlet valve, through a damper, into chamber 2, and out past the purge valve to the rest of the system. The two pistons are driven in opposite directions, and the first moves at twice the speed of the second.
Every stroke gives a small pressure pulse. Agilent’s 1260 specification allows below 2% or below 3 bar. When the pulse grows past that, the cause is usually in this module: a check valve, a seal, or air.
Point at a part in the drawing to open it.
The numbers
- Design (Agilent 1260)
- Dual piston in series: two chambers, one sapphire piston each 1
- Stroke volume, first piston
- 20 to 100 µL, set by the flow rate 1
- Flow range (Agilent 1260)
- 1 µL/min to 10 mL/min 1
- Pressure (Agilent 1260 quaternary)
- Up to 600 bar 1
- Pressure pulsation spec
- Below 2% amplitude, or below 3 bar 1
What goes wrong
- Pressure goes up and downCheck: Shimadzu's list: air bubbles, worn seals, check valves, leaks, poor degassing. A gradient changing viscosity is normal. 2
- Pump will not primeCheck: ASI's list: back pressure in the head (open the purge valve), a check valve fitted upside down, a clogged inlet filter or outlet valve filter 3
- Erratic pressure, and no leak you can seeCheck: Air leaking in at the inlet line. ASI: the intake side is under slight vacuum, so air leaks in and nothing drips out. 3
Module check
10 free questions on the parts of this module. Tap an answer, or a part of the drawing, to see if you are right and why.
Students get the full module test, with harder questions from real lab situations, and their scores saved. Students: take the module test. Not a student yet? See how to join.
Ready for the full module test? Harder questions from real lab situations, and your best score is saved. Take the module test →
Go deeper
- Pump types: isocratic, quaternary, binary One solvent, four solvents mixed before the pump, or two pumps mixed after. Where the mixing happens decides a lot.
- Pump head and pistons A steel head with a chamber, and a sapphire piston that moves in and out of it. The piston never touches the chamber wall; the seal does the sealing.
- Check valves One-way valves above and below the pump chamber. A ball sits on a seat and only lets solvent pass in one direction.
- Piston seals The ring that keeps the solvent in the chamber while the piston moves. It wears, and it is the most common pump repair.
- Seal wash A slow flow of wash solvent behind the piston seal, so buffer cannot dry there and form crystals.
- Purge valve A valve at the pump outlet that sends flow to waste, so you can prime the pump without going through the column. It has a frit that blocks.
- Inline filter A small filter on the high-pressure side, after the pump and before the column. It catches what the inlet filter cannot.
- Pressure and the damper The pressure reading is the most useful number on the instrument. Read the level, and read how steady it is.
Sources
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, pp. 19-22; How does variable stroke volume work?, p. 26; Specifications, PDF p. 42.
- Shimadzu, HPLC Troubleshooting Guide. Fluctuating pressure, p. 21.
- Analytical Scientific Instruments (ASI), Trouble Shooting Guide: pump replacement parts (white page). Erratic pressure and Failure to prime, pp. 1-3.