Pump
Inline filter
A small filter on the high-pressure side, after the pump and before the column. It catches what the inlet filter cannot.
The inlet filter in the bottle is 10 to 20 µm. The inline filter sits further down on the high-pressure side and catches the particles in between: from the mobile phase, from worn seals, and from salt that precipitated in a gradient valve.
It is a cheap part protecting an expensive one. When it loads up, the pressure rises; change the disc or the element, and the pressure comes back down.
Makers call it different names: Waters calls it the in-line filter, Thermo a pre-column filter. Waters warns not to replace it with a plain union, because the parts downstream lose their protection.
The numbers
- Where
- High-pressure side, between the pump and the sample manager or column 1
- Catches
- Particles from solvents, samples and worn parts such as seals 2
- Disc pore sizes (Agilent)
- 0.2, 0.5 or 2 µm 2
- Disc diameters (Agilent)
- 2.1 mm or 4.6 mm 2
- KNAUER cartridges
- 2 µm, in stainless steel, titanium (bio-inert) or PEEK 3
- Agilent gradient valve
- An inline filter can catch salt crystals from mixing and let them redissolve 4
- Flow direction
- Marked on KNAUER cartridges; fit it the right way round 3
Same part, different names
What goes wrong
Check yourself
Tap an answer to see if you are right and why.
Sources
- Waters Corporation, Alliance e2695 Separations Module Operator's Guide (715003794 Rev. B). Replace the in-line filter, PDF p. 196.
- Agilent Technologies, InfinityLab Quick Change Inline Filters for HPLC and UHPLC (product page, read 22 Nov 2026). InfinityLab Quick Change inline filters, product page: overview, Features, How it works.
- KNAUER, AZURA Pump P 6.1L User Manual (V6890). Maintenance and care, p. 57; spare parts list.
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Operational hints for the MCGV, p. 85.
- Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Section 4.6.2, Pre-column filters.