Pump
Purge valve
A valve at the pump outlet that sends flow to waste, so you can prime the pump without going through the column. It has a frit that blocks.
When you change a bottle or the pump has lost its prime, you want to flush the lines fast without sending air and old solvent through the column. The purge valve does that: open it, and the flow goes to waste.
Inside is a PTFE frit. It catches seal particles and anything else coming out of the pump heads, and over time it blocks.
Agilent gives a simple test: pump water at 5 mL/min with the purge valve open. More than 10 bar means the frit is blocked. Agilent also says to change it every time you change the seals.
The numbers
What goes wrong
Check yourself
Tap an answer to see if you are right and why.
Sources
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, p. 19; How does the pump work?, p. 22.
- Agilent Technologies, Agilent 1200 Series Quaternary Pump User Manual (G1311-90011). Operational hints, p. 50; Algae growth in HPLC systems, pp. 53-54.