Pump

Pump types: isocratic, quaternary, binary

One solvent, four solvents mixed before the pump, or two pumps mixed after. Where the mixing happens decides a lot.

PumpIsocraticone solvent, one channelABCDgradient valvePumpQuaternarymixes at low pressure, 1 pumpAPumpBPumpmixBinarytwo pumps, mixes at high pressure

Three designs cover most analytical pumps.

Isocratic. One solvent bottle, one line, one pump.

Quaternary, or low-pressure mixing. Up to 4 solvents go through a fast proportioning valve before the pump. Agilent calls it the MCGV. One pump does the work, and the order of the bottles matters (see Which solvent goes on which channel under Solvent lines).

Binary, or high-pressure mixing. Two pumps, one per solvent, and the solvents meet after the pumps. Agilent’s binary pump is two identical pumps in one housing.

Where the solvents meet decides the dwell volume, and the dwell volume is why a gradient method can behave differently on another instrument.

The numbers

Isocratic
One solvent line straight to the inlet valve 1
Quaternary (low-pressure mixing)
Up to 4 solvents; a fast proportioning valve mixes them before the pump 1
Binary (high-pressure mixing)
Two identical pumps in one housing; the solvents meet after the pumps 2
Degasser
Built in on Agilent's quaternary pump; a separate module for the binary pump 2

Same part, different names

Agilent
Multi channel gradient valve (MCGV) on the quaternary pump 1
KNAUER
LPG (low-pressure gradient) and HPG (high-pressure gradient) pump versions 3

What goes wrong

  • Gradient method moves when transferred to another instrumentCheck: Dwell volume. Low- and high-pressure mixing systems hold different volumes before the column. 4
  • Salt deposits or precipitation in the gradient valveCheck: Which channel the buffer is on. Agilent wants the salt solution on a bottom port. 1

Check yourself

Tap an answer to see if you are right and why.

Where are the solvents mixed in a quaternary (low-pressure mixing) pump?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, pp. 19-22; Operational hints for the MCGV, p. 85.
  2. Agilent Technologies, Agilent 1200 Series Binary Pump User Manual (G1312-90006). Introduction, PDF p. 6.
  3. KNAUER, 2021, Good HPLC Practice Guide (V6887). Mobile phase, p. 5, figure 1.
  4. Agilent Technologies, The LC Handbook. Dwell volume and its impact on chromatography, pp. 48-49.