Pump
Seal wash
A slow flow of wash solvent behind the piston seal, so buffer cannot dry there and form crystals.
Agilent says highly concentrated buffers shorten the life of the seals and pistons. The problem is buffer crystals on the back side of the seal.
Seal wash flushes that back side with a wash solvent. In Agilent’s version a small peristaltic pump moves the wash through the pump head and removes the buffer crystals. Agilent’s wash solution is 90% water and 10% isopropanol. The isopropanol keeps algae and bacteria out of the wash bottle.
On the Agilent 1260 seal wash is an option. Agilent’s rule for when you need it: 0.1 M buffer or more, used for long periods.
The numbers
- When (Agilent)
- Strongly recommended with buffers of 0.1 M or more over long periods 1
- Wash solution (Agilent)
- 90% water, 10% isopropanol, in a bottle above the pump 1
- Why that mix
- Stops algae and bacteria in the wash bottle and lowers the surface tension of water 1
- Agilent 1260
- Optional (active seal wash, peristaltic pump) 1
- Rinse solution (Shimadzu)
- 10% 2-propanol; replace once a week, even if it looks clean 2
What goes wrong
Check yourself
Tap an answer to see if you are right and why.
Sources
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). When to use the seal wash function, p. 86.