Pump

Pressure and the damper

The pressure reading is the most useful number on the instrument. Read the level, and read how steady it is.

Two things to read on the pressure display. The level, and how steady it is.

The level tells you about restriction. Same column, same flow, same mobile phase, higher pressure than last week: something is blocking. Shimadzu’s list of causes starts with the obvious (flow set too high) and then goes through the column, guard, precipitated buffer and the injector.

The steadiness tells you about the pump. Every stroke gives a small pulse, and the damper between the chambers smooths it. Agilent’s specification for the 1260 is below 2% or below 3 bar. A bigger, regular ripple points at a check valve, a seal or air; see the next page.

The numbers

Damper (Agilent)
A damping unit between the two chambers 1
Normal pulsation (Agilent 1260 spec)
Below 2% amplitude, or below 3 bar 1
Stroke volume
Smaller strokes give smaller, faster pulses 1
Too high, common causes
Flow set too high, blocked column or guard, precipitated buffer, blocked injector 2
Fluctuating, common causes
Air, worn seals, check valves, leaks, poor degassing 2

What goes wrong

  • Pressure higher than last week at the same conditionsCheck: Something is blocking: inline filter, guard, column frit, injector. Shimadzu lists the order. 2
  • Pressure cycles smoothly during a gradientCheck: Normal. Viscosity changes with composition. 2

Check yourself

Tap an answer to see if you are right and why.

What pressure pulsation does Agilent specify for the 1260 pump?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). How does the pump work?, p. 22; How does variable stroke volume work?, p. 26; Specifications, PDF p. 42.
  2. Shimadzu, HPLC Troubleshooting Guide. Fluctuating pressure, p. 21; High pressure, p. 22.