Pump
Pressure and the damper
The pressure reading is the most useful number on the instrument. Read the level, and read how steady it is.
Two things to read on the pressure display. The level, and how steady it is.
The level tells you about restriction. Same column, same flow, same mobile phase, higher pressure than last week: something is blocking. Shimadzu’s list of causes starts with the obvious (flow set too high) and then goes through the column, guard, precipitated buffer and the injector.
The steadiness tells you about the pump. Every stroke gives a small pulse, and the damper between the chambers smooths it. Agilent’s specification for the 1260 is below 2% or below 3 bar. A bigger, regular ripple points at a check valve, a seal or air; see the next page.
The numbers
- Damper (Agilent)
- A damping unit between the two chambers 1
- Normal pulsation (Agilent 1260 spec)
- Below 2% amplitude, or below 3 bar 1
- Stroke volume
- Smaller strokes give smaller, faster pulses 1
- Too high, common causes
- Flow set too high, blocked column or guard, precipitated buffer, blocked injector 2
- Fluctuating, common causes
- Air, worn seals, check valves, leaks, poor degassing 2
What goes wrong
Check yourself
Tap an answer to see if you are right and why.
Go deeper
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). How does the pump work?, p. 22; How does variable stroke volume work?, p. 26; Specifications, PDF p. 42.
- Shimadzu, HPLC Troubleshooting Guide. Fluctuating pressure, p. 21; High pressure, p. 22.