Solvent delivery
Degasser
Takes dissolved air out of the mobile phase before it reaches the pump.
Mobile phase holds dissolved air. When the pressure drops or the liquid warms up, the gas comes out as bubbles. A bubble in the pump head and the pump loses its prime. A bubble in the detector cell gives spikes and noise.
Most systems now have the degasser built in. Agilent’s handbook says that if it is bypassed, absent or not working, sparge with helium or degas some other way.
Agilent’s 1260 manual is direct about it: the built-in degasser should always be in the flow path.
The numbers
- Agilent 1260 quaternary pump
- Built-in vacuum degasser, 4 channels 1
- Low flow
- Agilent: below 500 µL/min a vacuum degasser is required 1
- After it was off
- Flush with at least 2 volumes (30 mL) before running, especially with volatile mixtures 2
- Autosampler wash liquid
- Thermo: degas it continuously too, using the pump's or the solvent rack's degasser 4
Same part, different names
What goes wrong
- The pump loses its primeCheck: Dissolved gas coming out of solution in the pump 3
- Spikes and noise on the baselineCheck: Gas coming out of solution in the detector cell 3
- Fluorescence or electrochemical response is low or driftingCheck: Dissolved air. Agilent says it quenches fluorescence and reacts at electrochemical detectors 3
Check yourself
Tap an answer to see if you are right and why.
Go deeper
- How a vacuum degasser worksStudents The solvent runs through gas-permeable tubing inside a vacuum chamber, and the gas crosses the membrane.
- Degassing off-lineStudents Helium sparging, sonication or vacuum filtration, when there is no online degasser or it is not working.
- When degassing matters mostStudents Low flow rates, aqueous and mixed solvents, temperature changes, and fluorescence or electrochemical detection.
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, p. 19; Specifications, PDF pp. 42-46; Using the degasser, p. 84.
- Agilent Technologies, Agilent 1200 Series Quaternary Pump User Manual (G1311-90011). Operational hints, p. 50.
- Agilent Technologies, The LC Handbook. Degassing mobile phases, PDF p. 73; Glossary, Degassing, PDF p. 149.
- Thermo Fisher Scientific, UltiMate 3000 Series SD, RS, BM and BX Pumps, Operating Instructions (4820.4001). Section 4.3, Solvent reservoirs, p. 58.
- IDEX Health & Science, In-Line Degassing Product Guide (Rev. A, 2023). Dissolved gas leads to bubble formation, p. 2.
- KNAUER, 2021, Good HPLC Practice Guide (V6887). How and why should the mobile phases be degassed?, p. 5.