Solvent delivery

Degasser

Takes dissolved air out of the mobile phase before it reaches the pump.

Mobile phase holds dissolved air. When the pressure drops or the liquid warms up, the gas comes out as bubbles. A bubble in the pump head and the pump loses its prime. A bubble in the detector cell gives spikes and noise.

Most systems now have the degasser built in. Agilent’s handbook says that if it is bypassed, absent or not working, sparge with helium or degas some other way.

Agilent’s 1260 manual is direct about it: the built-in degasser should always be in the flow path.

The numbers

Agilent 1260 quaternary pump
Built-in vacuum degasser, 4 channels 1
Low flow
Agilent: below 500 µL/min a vacuum degasser is required 1
After it was off
Flush with at least 2 volumes (30 mL) before running, especially with volatile mixtures 2
Autosampler wash liquid
Thermo: degas it continuously too, using the pump's or the solvent rack's degasser 4

Same part, different names

Agilent
Vacuum degasser (built in on quaternary pumps) 1
Thermo Scientific
Solvent Rack with integrated vacuum degasser (SRD-3x00) 4
KNAUER
Integrated module for online degassing 6

What goes wrong

  • The pump loses its primeCheck: Dissolved gas coming out of solution in the pump 3
  • Spikes and noise on the baselineCheck: Gas coming out of solution in the detector cell 3
  • Fluorescence or electrochemical response is low or driftingCheck: Dissolved air. Agilent says it quenches fluorescence and reacts at electrochemical detectors 3

Check yourself

Tap an answer to see if you are right and why.

Below which flow rate does Agilent say a vacuum degasser is required on the 1260 pump?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, p. 19; Specifications, PDF pp. 42-46; Using the degasser, p. 84.
  2. Agilent Technologies, Agilent 1200 Series Quaternary Pump User Manual (G1311-90011). Operational hints, p. 50.
  3. Agilent Technologies, The LC Handbook. Degassing mobile phases, PDF p. 73; Glossary, Degassing, PDF p. 149.
  4. Thermo Fisher Scientific, UltiMate 3000 Series SD, RS, BM and BX Pumps, Operating Instructions (4820.4001). Section 4.3, Solvent reservoirs, p. 58.
  5. IDEX Health & Science, In-Line Degassing Product Guide (Rev. A, 2023). Dissolved gas leads to bubble formation, p. 2.
  6. KNAUER, 2021, Good HPLC Practice Guide (V6887). How and why should the mobile phases be degassed?, p. 5.

New modules by email

All the modules are open now, and I keep adding parts and checks. Leave your email and I will send you what's new, including the Empower pages.

My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy