Solvent delivery
Solvent lines
Low-pressure PTFE tubing from each bottle to the degasser and the pump.
These lines run at low pressure, so the tubing is soft PTFE, 1/8 inch (3.2 mm) outside diameter. Agilent’s tube is 1.5 mm inside and 3 mm outside.
Thermo asks for the bottles close to the pump and the line to the degasser as short as possible.
After the pump, the tubing is 1/16 inch capillary, steel or PEEK. An Agilent 1260 quaternary pump can generate up to 600 bar. That is covered under Tubing and fittings.
The numbers
Same part, different names
What goes wrong
- Air in the line, degasser less effectiveCheck: A dirty connector. Thermo says minute particles can let air in. 3
Check yourself
Tap an answer to see if you are right and why.
Go deeper
The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.
Sources
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Parts: bottle head assembly, p. 183.
- Restek, 2021, Effective LC Troubleshooting: Symptom-Based Strategies and Solutions (GNAR3485), product pages. Bluestem glass solvent filter, product page, PDF p. 10.
- Thermo Fisher Scientific, UltiMate 3000 Series SD, RS, BM and BX Pumps, Operating Instructions (4820.4001). Section 4.3.1, General notes, p. 59.
- Agilent Technologies, Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual (G7111). Overview of the hydraulic path, p. 19 (600 bar).