Solvent delivery

Mobile phase in the bottle

What you put in the bottle decides how long the filters, check valves and seals last.

The inlet filter, the check valves and the pump seals all see what is in this bottle. So the mobile phase decides how long those parts last.

Three habits cover most of it: filter it, degas it, and make it fresh. Knauer says filtering helps the plungers, seals and check valves, and their lifetimes are longest that way.

How you mix also matters. Agilent’s example is 50/50 methanol and water: measure each volume on its own, then mix. Mixed in one container, you get a different composition.

The numbers

Filter buffers (HPLC)
0.45 µm 1
Filter buffers (UHPLC)
0.22 µm 1
Agilent pump manual
Always filter solvents through 0.4 µm filters 2
Buffers
Clear, no particles, prepared fresh on the day of use 3
Aqueous phase
Exchange every 2 days, or refilter 2

What goes wrong

  • Retention times move after the mobile phase was remadeCheck: How was it mixed? For 50/50 methanol/water, Agilent says to measure each volume separately before mixing, because the mixed volume is not the sum of the two. 1
  • Small particles block capillaries and valvesCheck: Unfiltered solvent. Agilent says small particles can block them permanently. 2

Check yourself

Tap an answer to see if you are right and why.

Your aqueous phosphate buffer has been in the bottle for a week. What does Agilent recommend?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, The LC Handbook. Mixing and degassing mobile phases, PDF p. 73; filtering buffers, PDF p. 77.
  2. Agilent Technologies, Agilent 1200 Series Quaternary Pump User Manual (G1311-90011). Solvent information, p. 51; Prevent blocking of solvent filters, p. 52.
  3. KNAUER, 2021, Good HPLC Practice Guide (V6887). Mobile phase, pp. 4-5.