sample tray · 4 °C vial needle seat wash port sample loop metering device injection valve

Point at a part, then click to open it. Back to the whole system

Sample manager

Sample tray and temperature

Holds the vials or plates, and on most systems can cool them so samples don't change while they wait.

Samples sit in the tray for hours in a long sequence. The tray is where you protect them.

Most systems can cool the tray. Agilent’s sample thermostat and Waters’ ACQUITY sample manager both hold 4 to 40 °C. Look at the small print, though: Waters gives that range for a room at 25 °C or less, and Thermo’s WPS-3000 cools by up to 22 °C below room temperature. On a hot day in a lab without good air conditioning, 4 °C may not be reachable.

When a long run shows the main peak shrinking while impurity peaks grow, think of the sample before the instrument. Shimadzu’s first fix is a fresh sample.

The numbers

Agilent vialsampler (sample thermostat option)
4 to 40 °C 1
Agilent 1260 vialsampler capacity
132 × 2 mL vials (two trays), or 36 × 6 mL vials 1
Waters ACQUITY sample manager
4 to 40 °C, in a room at 25 °C or less 2
Thermo WPS-3000 (thermostatted)
Can cool the samples by up to 22 °C below room temperature 3

What goes wrong

  • Main peak shrinks and impurity peaks grow across a long sequenceCheck: The sample is degrading while it waits. Shimadzu: prepare a fresh sample; cool the tray if the sample allows. 4
  • The tray will not reach 4 °C on a hot dayCheck: Waters specifies 4 °C only in a room at 25 °C or less; Thermo's cooler works relative to room temperature 2

Sources

  1. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Introduction, p. 9; Performance specifications, p. 27.
  2. Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). Sample manager, p. 1-7.
  3. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Sample thermostatting, PDF p. 28.
  4. Shimadzu, HPLC Troubleshooting Guide. Changes in sensitivity: sample, p. 16.