sample tray · 4 °C vial needle seat wash port sample loop metering device injection valve

Point at a part, then click to open it. Back to the whole system

Sample manager

Injection valve

A 2-position, 6-port rotary valve. In one position the pump bypasses the loop; turn it 60° and the pump pushes the loop contents onto the column.

LoadPump flow goes straight to the column.The needle fills the loop with sample.sample loopfrom pumpto columnneedlewasteInjectThe rotor turns 60°. The pump pushesthe sample out of the loop to the column.sample loopfrom pumpto columnneedlewaste

The autosamplers in this module, and the manual injection valves before them, are built around the same part: a 6-port rotary valve with two positions.

  • Load / bypass (left in the drawing): the pump flow goes straight from one port to the next and on to the column. The loop is off line, so the needle can fill it with sample (yellow).
  • Inject / mainpass (right): the rotor turns 60°. The orange grooves now join different ports, so the pump flow runs through the loop and carries the sample to the column.

Inside, a rotor seal turns against a fixed stator, and grooves in the rotor connect pairs of ports. Thermo lists both the rotor seal and the stator face seal as consumables: parts you replace, like piston seals in the pump.

The vendors don’t use the same words for the two positions. Agilent says mainpass and bypass; Thermo and IDEX say Inject and Load.

The numbers

Type
2-position, 6-port 1
Switching
A 60° turn between Load and Inject (IDEX manual valve) 2
Wear parts
Rotor seal and stator face seal are consumables 1
Port-to-port volume (IDEX V-450, manual)
5.1 µL 2
Do not
Use concentrated formic acid; let water freeze in the valve below 0 °C (Thermo) 1

Same part, different names

Agilent
Injection valve: mainpass and bypass positions 3
Thermo Scientific
Injection valve: Inject and Load positions 1
IDEX
Injection valve: LOAD and INJECT 2

What goes wrong

  • Leak at the valve, or the pressure drops at each injectionCheck: A worn rotor seal or stator face seal. Replace them as a pair of consumables. 1
  • Leak after fitting new capillaries to the valveCheck: Thermo warns against mixing ferrule types on its valve; use the fittings made for that valve. 1

Check yourself

Tap an answer to see if you are right and why.

How many ports does a standard autosampler injection valve have?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Injection valve, PDF p. 155 (manual p. 147).
  2. IDEX Health & Science, 6-Port Medium Pressure Injection Valve Instruction Guide (web page, read 22 Nov 2026). What is the medium pressure injection valve; Specifications.
  3. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Injection sequence, pp. 16-17.