Sample manager

Metering device and syringe

The small piston pump or syringe that draws the sample. Its stroke sets the injection volume.

The metering device is what actually measures your injection. It is a small piston pump (Agilent calls it the analytical head, Shimadzu the measuring pump) or a syringe (Waters, Thermo). Its stroke pulls the set volume of sample through the needle into the loop.

So when areas scatter but retention times are fine, this is one of the first places to look. Thermo’s list for small peaks includes a cracked or dirty syringe and a sticking plunger, often from particles in the sample. Viscous samples need a longer draw time, or less is drawn than you set.

Agilent’s metering device has its own wear parts, a metering seal and piston, like a small version of the pump head.

The numbers

Agilent vialsampler
Metering device (analytical head) draws the sample into the loop; 0.7 µL air plug after drawing 1
Agilent analytical heads
40 µL, 100 µL and 900 µL versions 1
Thermo WPS-3000
Syringe draws the sample and also supplies the wash port; a buffer loop keeps mobile phase out of the syringe 2
Waters ACQUITY
Sample syringe pulls the sample through the needle and valve 3
Wear parts (Agilent)
Metering seal and piston 1

Same part, different names

Agilent
Metering device, analytical head 1
Waters
Sample syringe 3
Thermo Scientific
Syringe, buffer loop 2
Shimadzu
Measuring pump 4

What goes wrong

  • Peaks smaller than expected, or areas scatterCheck: Thermo: a cracked or dirty syringe, or a sticking plunger, often from particles in the sample 5
  • Areas low or variable with a viscous sampleCheck: The draw time is too short for the viscosity, so less is drawn than set (Thermo) 5
  • Sensitivity changes from run to runCheck: Shimadzu's list for an injector cause: check the dispensed volume with a system suitability sample, and purge the injector metering pump 6

Check yourself

Tap an answer to see if you are right and why.

What sets the injection volume in most modern autosamplers?

Sources

  1. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Sampling sequence and injection sequence, pp. 15-18; parts list (analytical heads); Exchange the metering seal and piston.
  2. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Operating principle, PDF pp. 24-25.
  3. Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). How sample flows, p. 1-7.
  4. Shimadzu, SIL-30AC Autosampler Instruction Manual. Sample injection flow diagram, p. 18.
  5. Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Smaller than expected peaks, PDF p. 19.
  6. Shimadzu, HPLC Troubleshooting Guide. Changes in sensitivity: injector issue, p. 16.