Sample manager
Sample loop
The piece of tubing that holds the sample until the valve puts it in the flow. Its size limits how much you can inject.
The loop is a coil of tubing on the injection valve, or in a flow-through design the tubing between the needle and the valve. It holds the sample while the pump flow bypasses it, then the valve puts it in line.
Its size sets your upper limit. On Agilent’s 1260 vialsampler the standard head handles 0.1 to 100 µL; bigger volumes need a different analytical head or multiple draws. Waters adds extension loops to its flow-through needle.
The loop is rarely the real limit, though. The column is. Thermo’s rule of thumb: about 10% of the column void volume if the sample is in a weaker solvent than the mobile phase, but only about 1% if it is in a stronger one. Past that, early peaks broaden.
The numbers
- Agilent 1260 vialsampler
- 100 µL standard; 900 µL analytical head for up to 900 µL; up to 1800 µL with multiple draws 1
- Shimadzu SIL-30AC
- Sample loop 0.3 mm × 1200 mm, 84.8 µL 2
- Waters SM-FTN
- Extension loops between needle and valve increase the injection volume 3
- Partial loop limit
- Inject no more than half the loop volume 4
- Injection volume vs column (Thermo)
- Up to about 10% of the column void volume in a weak solvent; only about 1% in a strong solvent 5
What goes wrong
- Early peaks broad or split after you raised the injection volumeCheck: Too much volume, or sample in a solvent stronger than the mobile phase. Thermo: about 10% of void volume in a weak solvent, 1% in a strong one. 5
- Peaks smaller than expectedCheck: The wrong loop is fitted. Thermo puts this on its list for small peaks. 5
Check yourself
Tap an answer to see if you are right and why.
Sources
- Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Performance specifications, p. 27.
- Shimadzu, SIL-30AC Autosampler Instruction Manual. Holding capacity in the flow line, p. 18.
- Waters, ACQUITY UPLC Sample Manager - Flow Through Needle: Operator's Overview and Maintenance Information (Rev C). Overview, p. 1.
- IDEX Health & Science, 6-Port Medium Pressure Injection Valve Instruction Guide (web page, read 22 Nov 2026). Partial loop filling.
- Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Early eluting peaks broad, PDF p. 20; Smaller than expected peaks, PDF p. 19.