sample tray · 4 °C vial needle seat wash port sample loop metering device injection valve

Point at a part, then click to open it. Back to the whole system

Sample manager

Needle and needle seat

The needle draws the sample; in flow-through designs it then sits in a seat that seals it against full system pressure.

The needle goes into the vial and draws the sample. In a flow-through design it then goes down into the needle seat, and the seat has to seal it against the full pressure of the pump. Thermo says the seat’s shape is also there to avoid carryover and keep air out.

That makes the seat one of the main places carryover hides. Agilent’s fix when carryover is far higher than it should be: pipette a few millilitres of a solvent that dissolves the contamination onto the seat, wipe it, and if you don’t know what the contamination is, use two or three solvents of different polarity.

A new needle or seat is not at its best straight away. Agilent’s 1200 HiP manual says it can take 100 to 200 injections to reach the lowest carryover.

Waters’ ACQUITY uses a puncture needle to pierce the septum or plate cover, and the thinner sample needle comes out of it to draw the sample.

The numbers

Needle seat (Thermo)
Seals the needle tip against the system pressure; its shape avoids carryover and keeps air out 1
Needle height (Agilent vialsampler)
Default 2 mm above the vial bottom; adjust the offset for microvials 2
Minimum sample (same)
1 µL from 5 µL in a 100 µL microvial 2
After a new needle or seat (Agilent 1200 HiP)
It may take 100 to 200 injections to reach the best carryover 3
Shimadzu SIL-30AC needle volume
11.7 µL 5

Same part, different names

Agilent
Needle, needle seat assembly 2
Waters
Sample needle, puncture needle, needle seal 4
Thermo Scientific
Needle, needle seat, needle seat capillary 1

What goes wrong

  • Carryover stays high after you fixed the washCheck: Clean the needle seat with a solvent that dissolves the contamination, or replace it (Agilent) 2
  • Carryover goes up after work on the seatCheck: Air bubbles in the needle seat connection can raise carryover for a while; purge them out (Thermo) 1
  • No peak, or very small peaksCheck: A blocked needle. Shimadzu adds: find out why it blocked, for example septa or poor sample preparation. 6

Check yourself

Tap an answer to see if you are right and why.

In a flow-through needle design, what does the needle seat do?

Sources

  1. Thermo Fisher Scientific, UltiMate 3000 Series WPS-3000 Autosamplers, Operating Instructions (4828.2250). Operating principle, PDF p. 24; needle seat tip, PDF p. 154.
  2. Agilent Technologies, InfinityLab Vialsamplers User Manual (SD-29000242). Performance specifications, p. 27; General recommendation for lowest carryover, p. 187.
  3. Agilent Technologies, 1200 Series High Performance Autosampler User Manual (G1367-90011). Needle and seat replacement and carryover, p. 68.
  4. Waters Corporation, ACQUITY UPLC System Operator's Guide (71500082502 Rev. F). How sample flows, p. 1-7.
  5. Shimadzu, SIL-30AC Autosampler Instruction Manual. Holding capacity in the flow line, p. 18.
  6. Shimadzu, HPLC Troubleshooting Guide. No peaks: blocked needle, p. 8.